Monitor Setup
VF4 Infinity Gamma Series Page 2-19
Turning External Alarm Lights ON/OFF
A set of alarm lights on top of the monitor blink red for life-
threatening alarms and yellow for serious alarms, if the external
alarm light function is enabled in the Monitor Setup menu
(default setting is ON). If more than one alarm occurs at the same
time, the lights blink for the alarm with the highest alarm grade. If
the user silences the alarm, the alarm lights remain lit without
flashing. If the user turns all alarms off, the alarm lights are
turned off as well.
STEPS: Turning External Alarm Lights On/Off
1. Press the Menu fixed key.
2. Click on
Monitor Setup.
3. Click on
Monitor Options.
4. Click on
Alarm Light.
5. Select
ON or OFF and click the knob again.
WARN IN G: The alarm light setting (ON/OFF) is not indi-
cated on the screen. Before monitoring your patient, ver-
ify the alarm light setting in the Alarm Light menu.
NOTES:
! The external alarm lights do not flash for advisory alarms.
! During startup, the alarm lights blink briefly as part of the
monitor’s functional check.