Page 10-8 Infinity Gamma Series VF4
ST Alarms
Set ST alarm limits on the Alarm Limits table (see the chapter
Alarms and Messages). The default ST segment deviation alarm
limits are +1.0 mm (or +0.10 mV) for the upper limit and -1.0
mm (or -0.10 mV) for the lower limit. However, adjustments to
the upper and lower alarm limit settings must be made based on
careful clinical evaluation. You can set the upper and lower lim-
its within the following range:
! -15.0 mm to +15.0 mm, in 0.1 mm increments (-1.5 mV to
+1.5 mV, in 0.01 mV increments, when using millivolts.)
NOTE: Always check the upper and lower ST alarm limits before
starting ST monitoring.