Page 5-10 Infinity Gamma Series VF4
Alarm Validation
The alarm validation feature minimizes nuisance alarms for tran-
sient conditions. When a parameter value falls below or rises
above its current alarm limits, the monitor waits a predetermined
time before triggering the alarm. If the parameter value returns to
within the upper or lower alarm limit before the end of the alarm
validation period, the monitor does not trigger the alarm.
The alarm validation period is part of the software (see table
below) and cannot be modified. The monitor triggers the alarm at
the end of the validation period. Parameters have the following
alarm validation periods:
Parameter-Specific Alarm Validation Periods
Parameter
Upper Limit
(in seconds)
Lower Limit
(in seconds)
ECG/Heart rate (HR) 2 0
Pulse rate (PLS) 4 10
ST Segment Analysis* 60 60
Respiration rate (Rsp) 8 10
Arterial Oxygen Saturation (SpO
2
)4 10
etCO
2
*, iCO
2
*, T, PVC, Arrhythmia
Events
00
Scio Multigas* 0 0
RRc* 8 10
Invasive blood pressure (ART, PA,
CVP, ICP, GP1, GP2*)
44
Non-invasive blood pressure (NBP) 0 0
*Available only if option is enabled

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Gamma Series and is the answer not in the manual?

Dräger Infinity Gamma Series Specifications

General IconGeneral
BrandDräger
ModelInfinity Gamma Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals