Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 525
Troubleshooting
! %0 lead off
1), 2)
The parameter
value is replaced
by ***
The RRi lead is invalid.
– Check the
patient and
treat if neces-
sary.
– Make sure the
patient’s skin is
prepared prop-
erly.
– Isolate the pa-
tient from any
auxiliary equip-
ment, if possi-
ble.
– Reapply gel on
reusable elec-
trodes and re-
apply them or
replace new
disposable
electrodes.
– Inspect and
replace faulty
cables and
wires.
– If lead or
electrode
cannot be
replaced, se-
lect another
lead for
processing (in
the RRi setup
page).
! %0 artifact
1), 2)
The parameter
value is replaced
by ***
Persistent artifact was
detected.
!
RRi high
impedance
The parameter
value is replaced
by ***
A high respiration
impedance was
detected.
!
%0 lead
unavailable
1)
The parameter
value is replaced
by ***
Faulty or disconnected
electrodes.
1)%0 is a placeholder for the parameter label RRi.
2) After discharging a patient or starting the device, the alarm becomes active only after a numeric value
has been received for that parameter.
Priority Message Parameter field Problem Solution
NOTE
RRi and 12-lead ECG monitoring are unavailable
when the M540 is in OR mode and the ECG filter
is set to Monitor.