Troubleshooting
532 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
None
(Any
SpO2
MCable)
SpO2 searching Searching
Parameter values
are replaced by ***
The sensor is
searching for valid
pulses to compute a
measurement value.
Verify proper
sensor application.
!
(Any
Masimo
MCable)
SpO2 sensor
expired
3)
Replace sensor
Parameter values
are replaced by ***
–SpO2 sensor
expired
Adhesive sensor
expired
Replace the
sensor.
!
(Any
Masimo
MCable)
SpO2 sensor expires
soon
Sensor expires
soon
Parameter values
–SpO2 sensor near
expiration
Adhesive sensor
near expiration
Replace the
sensor.
!
(Any
SpO2
MCable)
%0 sensor failure
1)
Sensor failure
Parameter values
are replaced by ***
Hardware failure
Faulty SpO2
sensor
–Make sure the
SpO2 sensor is
properly
attached to the
patient and all
cables are
properly
connected.
Replace the
sensor.
Contact
Dräger-
authorized
service
personnel.
! SpO2 sensor off
2)
Sensor off
Parameter values
are replaced by ***
The Masimo MCable
has detected that the
SpO2 sensor is no
longer attached to the
patient.
Reattach the SpO2
sensor.
Priority Message Parameter field Problem Solution