Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 545
Troubleshooting
!! CO
2 warming up The CO2 sensor is
completing its warm-up
cycle.
Wait for the CO2 sensor to
warm up. During warm-up,
the accuracy is reduced.
If the message persists
longer than 15 minutes
after the sensor has
warmed up and the ambient
temperature is above 10 °C
(50 °F), contact
DrägerService
®
.
You cannot zero the sensor
while this message is
displayed and the ambient
temperature is above 10 °C
(50 °F).
When the ambient
temperature is below 10 °C
(50 °F), the message can
display longer than
15 minutes. In this case, it
is possible to zero the
sensor after the message
has been displayed for at
least 10 minutes.
None %0 zeroing failed Zeroing of the sensor has
failed or the sensor is faulty .
Try to zero the sensor again
making sure not to breathe
on the sensor.
If zeroing fails again,
replace the sensor and
contact DrägerService
®
.
None %0 zeroing in progress The CO
2 zeroing is in
progress.
Informational message – no
action required.
!! etCO
2 > (alarm limit)
etCO
2 < (alarm limit)
(except inCO2)
The parameter value is
above/below the set
upper/lower alarm limits.
Check the patient and treat
if necessary.
Change the alarm limits.
!! RRc out of range high The parameter signal is
outside the measuring
range of the monitor.
Check the patient and treat
if necessary.
Check the equipment and
replace if necessary.