Technical data
598 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the equipment is used exceeds the applicable RF compliance level above, the
equipment should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the equipment.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances
Recommended separation distances between portable and mobile RF telecommunication
devices and the Cockpit
max.
P
EIRP (W)
150 kHz to
800 MHz
Distance
1)
(m)
800 MHz to
2.5 GHz
Distance
1)
(m)
Comments (if applicable)
0.001 0.04 0.07
0.003 0.06 0.12
0.010 0.12 0.23
0.040 0.21 0.4 For example: WLAN 5250
0.100 0.38 0.73 For example: WLAN 2440 (Europe), Bluetooth
0.200 0.54 1.03 For example: WLAN 5250 (Europe)
0.250 0.6 1.03 For example: DECT-devices
1.000 1.2 2.3 For example: GSM 1800 / GSM 1900 / UMTS cellular
phones, WLAN 5600 (not in Europe)
2.000 1.7 3.25 For example: GSM 900 cellular phones
3.000 2.08 3.98
10.00 3.8 7.27
100.00 12 23
1)
NOTE: Information regarding separation distances (IEC 60601-1-2:2007, tables 4 and 6)
Electromagnetic immunity
Immunity
against...
IEC 60601-1-2 test level Compliance
level (of device)
Electromagnetic environment