Disassembly/Assembly Flow Sensor
Open the flap by lifting the
lower edge upwards.
Close the flap.
Pull the flow sensor as far
as it will go to the left.
Push the flow sensor to
the right into the socket.
Remove the expiratory valve
from the fitting.
Turn the locking ring as far as
possible to the left.
Remove and empty the water
trap container.
Fit the water trap container.
Remove the diaphragm.
Fit the flow sensor
into the socket.
Remove the flow sensor
from the socket.
Pull the sensor sleeve from
the expiratory valve.
Fit the diaphragm onto the edge
of the expiratory valve housing.
Fit the flow sensor sleeve
on the expiratory valve.
Turn the locking ring as far
as possible to the right
until it clicks into place.
Turn the locking ring as far as
possible to the left, push the
expiratory valve into the fitting.
Dräger Savina
®
300/Classic/Select
Disassembly/Assembly
Disassembly/Assembly Expiratory Valve
Dräger Savina
®
300 Classic
D-12306-2016
Dräger Savina
®
300 Select
D-12295-2016
Dräger Savina
®
300
D-64714-2012