4. In case of needing to replace the fuse, the power should be cut off first, followed by the
opening of cover to replace the fuse.
5. When heating the water tank, do not touch with hand to prevent the scald.
6. Can not be used under the environment full of flammable, explosive and narcotic gas.
7. In general, this product does not have specific requirements for disinfection.
IV. Main technical parameters
Temperature
inside water
tank
V. Maintenance
1. Glass component: in case of serious dirt, clean it with the dilute sulfuric acid.
2. PTFE connection tube: in general, it needs to be removed and washed only under serious
dirt.
3. The clear water inside the water tank pot should be frequently replaced, with the pot cleaned.
VI. Failure analysis and exclusion methods
1. Check whether the power plug is loose
2. Check whether the fuse is disconnected
1. Insert the power plug
tight
2. Replace the fuse
Rotation
bottle
shakes or gives
out noise
Check whether the glass interface of rotation
bottle is installed smoothly and in place
Reinstall the glass interface
and rotation bottle, and
screw and lock the nut
tightly.
1. Check the performance of vacuum pump, and
whether each connector is tightened (block the
air outlet of vacuum pump to observe the
indicators on vacuum table)
2. Check whether the glass appliances
are defected or cracked for coldness
3. Check whether the interface of glass appliance
is
coated with vacuum
1. Screw the
connectors tightly or
replace the vacuum pump
2.Replace the
damaged
appliances
3. Recoat with
vacuum grease
Water tank
can not
be heated.
1 Check the 10A fuse
2. Check whether the heating plug is loose
1. Replace the 10A fuse
2. Insert the heating
plug tightly
In case of being unable to troubleshoot according to the above methods, contact us by calling