EasyManua.ls Logo

Drive MQ5800 - Specifications; Compressor and Nebulizer Specs; Equipment Classifications

Drive MQ5800
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
.............................................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL MQ5800, MQ5900, MQ5900P
.............................................................................................................................
13
CAUTION
WARNING!
Size 7.5 in X 3.7 in X 5.7 in
Weight 3.0 lbs
Operating Voltage 120VAC, single phase, 60Hz
Current Consumption 2A (Max)
Storage/Transport Temperature Range -13 to 158°F, (-25 to 70°C)
Storage/Transport Humidity 10 – 95%
Operating Humidity 10 – 95%
Operating Temperature Range 50 to 104°F, (10 to 40°C)
Compressor Flow Rate ≥9 L/minute
Operating Pressure Range 8~13 psi /55Kpa~90Kpa
Operating Noise Level 52 +\-5 dB(A)
Service Life 1080 hours
18260 JetNeb Nebulizer
Min & Max lling volume 2~8 ml
Particle size (MMAD) <5 µm
Nebulization rate ≥0.25 ml/min.
Disposable nebulizer 30 days
For specification of NEBKIT 500: refer to the user manual of NEBKIT 500, which is
included in the accessory pack of NEBKIT 500 independently.
Equipment Classifications:
Protection against electric shock: Class II
Degree of protection against electric shock: Type BF
Equipment not suitable for use in the presence of ammable anaesthetic mixture with air.
Not intended for use with oxygen enriched environments such as oxygen hoods or tents.
*Conditions may vary based on altitude above sea level,barometric pressure, and temperature.
**Value determined with minimum ow rate through the nebulizer (20°C, sea level).
There are no hazardous materials used in the construction of this unit, therefore, there are
no known risks associated with disposal of the equipment.
This product complies with the following electromagnetic compatibility standard:
IEC 60601-1-2
Please dispose of this nebulizer in accordance with local, municipal, state
and federal law for electronic devices
specications