49
RMO60TD Accessories
9 Technical Data
9.1 Mains Power Supply
- Connection according to IEC/EN60320-1; CSA 22.2
- Voltage 90 V 264 V AC, 50 / 60 Hz, single-phase
- Input power 2250 VA
- Fuse 15 A / 250 V, type F
9.2 Output data
- Test current 5 mA 60 A DC
- Measuring range / Resolution
0,1 µΩ - 999,9 µΩ 0,1 µΩ
1,000 m- 9,999 m 1 µΩ
10,00 m- 99,99 m 10 µΩ
100,0 m- 999,9 m 0,1 m
1,000 Ω - 99,99 Ω 10 mΩ
100,0 Ω - 999,9 Ω 0,1 Ω
1000 Ω - 2000 Ω 1 Ω
- Typical Accuracy ± (0,1 % rdg + 0,1 % FS)
9.3 AC Current monitoring channel
- Current monitoring resolution: 0,1 ms
- Amplitude resolution: 16 bit
9.4 Environmental conditions
- Operating temperature -10 °C - +55 °C / 14 F - 131 F
- Storage temperature -25°C - +70 °C / -13 F - 158 F
- Relative humidity 5 % - 95 %, non-condensing
9.5 Dimensions and Weight
- Dimensions 450 mm x 175 mm x 320 mm (W x H x D)
17,72 in x 6,89 in x 12,6 in
- Weight 13 kg / 28,6 lb
9.6 Warranty
three years
9.7 Applicable Standards
- Installation/overvoltage: category II
- Pollution: degree 2
- Safety LVD 2006/95/EC, (CE Conform)
Standard EN 61010-1
- EMC Directive 2004/108/EC (CE Conform)
Standard EN 61326:2006
9.8 USB Interface
RMO60TD is equipped with an USB serial interface to connect to an external computer.