Dynisco ATC 990 User Manual

Default Icon
13 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #4 background imageLoading...
Page #4 background image
!
!
A347$#(S76;9(A9>$G/ 63'>73$&(178$U473$ '#(
!
If!e nabled,!the!semi‐automatic!calibration!uses!a!resistor!present!with in!the!pres sure!transducer,!to!calibrate!the!instrument!
to!the! transducer .!The!resistor!i s!used!to!force!the!mV!output!from!tra nsd ucer!to!a!fi xed!percentage!of!the!full!scale!value!
wh en!the!calibration !relay!shorts!out!the!R‐CAL!connections.!Strain!Gauge!Ca libration!is!a!s ub‐menu!option!i n!Input!
Configuration,!but!can!also!be!accesse d!at!any!time!by!pressing!DOWN!and!FORWARD!s imultaneously.!Allow !time!for!the!
instrument!and!process! to!reach!operating!temperat ure!before!starting.!The!first!strain!gauge!input!is!calibrated!the n!the!
sec ond!if!i t!is!fitted!and!con figured .!
1. Select!the!Shunt!Resistor!as!K#7U895.!If#enabled#the#semi‐automatic#calibration#procedure#is#used.#If#it#is#disabled,#use#
the#Manual#calibration#i nstructions#below .!
2. Set !the!Calibration!Resis tor!value!(default!=!80%!of!sensor!maximu m).!The#instrument#must#know#wha t#percentage#will#
be#applied#to# calibrate#correctly#–#Check#sensor#datasheet.!
3. At!the! Low!Po int!Cali bration!screen,!ensure!the!tr ansduc er!is!at!workin g!te mperat ure!with!zero!pressure!applied,!then!
press!UP!and!DOWN!simultaneously.!Low#end#calibratio n#is#per formed #and#a#completed#message#shown.#The#error#
message#“Count#Failure”#is#shown#if#the#input#was#not#± 10mV#of#the#nominal#0m V#expected#sig nal,#and#calibration#is#not#
altered.#0J$%(&'685(%$;#$X"(7(X768 3"(347#%56&94Y (Z'6(&7#('#8"(>'[9(' #3'(J$;J(9#5(&78$U473$'#('#&9(3J9(<4'U8 9>($%(
&'449& 395(7#5(8'\(&78$U473$'#(J 7%(U99#(&' ><8939 5(%6&&9%%X688"]!
4. At!the! High!Point!Cal ibration!screen,!press!UP!and!DOWN!simultaneously. !The!calibration!relay!turns!ON!to!short!the!R‐
CAL!terminals!together!and!force!required!%mV!value!to!be !output!from!transdu cer!into!the!instruments! input.!High#
end#calibration#is#performed#and#a#completed#message#shown.#The#err or#message#“R #Cal#Failure”#is#shown#if#the#signal#
was#less!than!+20mV!or!greater!than!+ 50mV,#and#calibration#is#not#altered.#0J$%(&'68 5(%$;#$X "(7(X768 3"(347#%56&94(+G178(
49% $%3' 4Y(+9<973(3 J9(&78$U473$'#( <4'&95 649('#&9(3J9(<4'U89>($%(&'449&395Y!
5. The!procedure!from!1!to!4!is!repeated!if!a!2
nd
!str ain!gauge!input!is !configured.!
!
!
A347$#(S76;9(^7#678(178$U473$'#(
!
If!the!shunt!is!disabled,!a!Manual!Calibration!needs!to!be!p erformed!by!user. !Manual!Calibration!requires!the!user!t o!input!
known!Zero!Offset!and!Full! Scale!mV!values!into! the!inp ut!terminals!of!the!in strument!using!an!accurate!mV!calibration!
source .!
!
The!Manu al!Calibration! process!for!the!low!offse t!is!as!follows: !!
1. Ensure!Unit!is!power ed!on!and!mV!Input!from!th e!calibration!signal!source!is !wired!to!the!instrum ent’s!st rain!gauge!
input!terminals.!
2. Enter!the!Strain!Gauge!Calibration!mode,!(e.g.!pr ess! DOWN!and!FORWARD!simul taneou sly).!
3. Select!the!Shunt!Resistor!as!Disabled.!
4. The!Calibr ation!R esistor!percentage!value! is!fi xed!to!100.0%!and!ca nnot!be!changed!by!t he!user.!
5. At!the! Low!Po int!Cali bration!screen,!check!that!the!calibration!signal!is!set!to! exactly!the!tra nsducer!mV!signal!at!zero!
pressu re,!then!press!UP!and!DOWN!simultaneously.!Low#end#calibration#is#pe rfor med#and#a#completed#message#shown.#
The#error#message#“Count#Failure”#is#shown#if#the#input#was#no t#±10mV#of#the#nominal#0mV#expected#signal,# and#
calibration#is# not#altered.#1J9&F("' 64(% $;#78(7#5(49<973(3J9(&78$U473 $'#( <4'&9%%(X4'>(3J9(%3743Y(Z'6(&7#('#8"(> '[9('#3'(
J$;J(9#5(&78$U473$ '#('#&9(8'\(&78$U473$'#(J7%(U99#(&'><89395(% 6&&9%%X688"]!
6. At!the! High!Point!Cal ibration!screen,!check!that!the!c alibrati on!signal!is!set!to!the!full!scale!mV! signal!as!per!the!
transducer!datasheets!(e.g.! 33.3mV),!then !press!UP!and!DOWN!simultaneously.!High#end#calibration#is#pe rfor med#and#a#
completed#message#shown.#The#error#message#“R#Ca l#Fa ilure”#is#shown#and#calibration#i s#not#altered#if#the#signal#was#
less!than!+20mV!or!greater!than!+50mV!O R!the!High!Cal!–!Low!Cal! value!is!less!than!10mV.#1J9&F("'64(%$;#78(7#5(
49<973(3J9(&78$U473$'#(<4'&9%%(X4'>(3J 9(%3 743Y !

Questions and Answers: