EasyManua.ls Logo

Dynisco ATC 990 - Page 5

Default Icon
13 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
!
!
!
178$U473$'#(+9>$#594!
!
A!calib ration!reminder!can!be!set!if!Da ta!Recorder!option!is!fitted!( which!means!instrument!has!battery‐b acked!Real!Time!
Clock).!The!addition!of!a!clock!means!that!the!instru ment!can!be!a war e!of!the!date!and! time!(even!wh en!powered!down).!
This!in formation! can!be! used!to!remind!the!user!to!c alibrate!the!inputs!at!a!specified!future!date.!The!date!is !set!in!the!Input!
Configuration!menu.!If!enabled!and!the!date!set!is!exceeded,!a!calibration!due!screen!is!shown!every!24h rs!(and!at!ever y!
power‐up)!until!it!is!change d!to!a!future!date!(or!disabled).!
!
V$#97 4(7#5(09><947364 9(178$U473$'#(.(A&78$#;(
)*+,--(.(/01, , -!
V$#974(M#'#G%347$#(;76;9R(7#5(09><9473649(178$U473$'#(
(
The!instru ment!o ffers!single!point!calibration ,!with!z ero!offset!adjustment!only,!or!two!point!calibration,! with!both!zero!&!
span!adjustments.!
!
A$#;89(*'$#3(178$U473$'#(
!
A!‘zero!off set’!applied!to!the!process!varia ble! across!the! entire!span.!Positive!values!are!add ed!to!the!reading, !negative!
values!are!subtracted.!Can!be!used !if!the!error!is!con stant!across!the!range,!or!the!user!is!only!interested!i n!a! single!critical!
value.!
To! use,!select!Single!Point!C alib ration!from!th e!input !calibration!menu,!and!enter!a!value!equal,!but!oppo site !to!the!obs erved!
error!to!correct!the!reading. !
!
!
!
This!example!shows! a!positive!offset!value.!
For!example:!
If!the!process!displays!27.8!when!it!should!
rea d!30,!The!error!is!‐2.2 !so!an!applied!
offset!of!+2.2!would!change!the!displayed!
value!to!30.!
The!same!offset!is!applied!to!al l!values,!so!
at!100.0!the!new!displayed!value!would!be!
102.2.!
!
(
0\ '(*'$#3(178$U473$'#(
!
This!m ethod!i s!used!where!an!error!is!not!constant!across!the!rang e.!
Separate!offsets!are!app lied!at!two!points !in!the!range!to!eliminate!both!“zero”!and!“span”!err ors.!To!use:!
1. Measure!and! record!the!error!at!a!l ow! point!in!th e!process.!
2. Measure!and! record!the!error!at!a!high!po int!in!the!proc ess.!
3. Go!to!the!first!two!point!input!calibration !screen.!
a. Enter!the!desired !low!point!value!as!the!Calibrati on!Low!PV!value.!
Single!Point!‘Offset!
Calibration’!value!
!
Original!Displayed!
Value!

Related product manuals