User ManualEW-M48
16
1
2
3
4
5
6
7
The maximum charging current for the LiNX LE System is 6A.
The wheelchair manufacturer must specify an appropriate baery charger for the baeries used in the
wheelchair.
The wheelchair manufacturer must also specify the maximum current of any baery chargers to be used with the
controller and warn against using baery chargers of higher current rangs.
The baery charger must have over-current protecon in the form of a non-reseable fuse.
It is the responsibility of the wheelchair manufacturer to manage the risks of baery over-charging and any
related gas emissions.
To protect the wheelchair wiring from over currents while charging the baeries, chargers must have the ability
to reduce their current output when electrically shorted.