Repetitive Strength. This module records the number of times that a user achieves tongue pressure within
a desired range. Target ranges should be determined by your SLP. To perform the exercise:
1. Open the Repetitive Strength module within the Tongueometer application.
2. Adjust the target range by sliding the blue dots on the “Min-Max Bar” to set the desired
minimum and maximum pressure range.
o If you swipe to another module or close and reopen the application, the pressure range
on the Repetitive Strength module will not change.
o The application requires there to be at least a 5 kPa difference between the minimum
and maximum setting.
3. Place the bulb on your tongue and squeeze it against the roof of your mouth until the meter’s
needle moves into the desired range highlighted in blue.
4. Once the needle moves into the highlighted range, stop applying pressure to the bulb. The
bulb may remain in the mouth between repetitions.
o The needle must return to 0 before another repetition will be counted.
o The counter at the bottom of the screen keeps track of each successful repetition during
which the pressure applied to the bulb enters the highlighted range.
o The RESET button resets the counter to zero. The Data Report will store practice
information even if a user presses RESET.