The following information recommends certain boundaries that will allow the
ECLIPSE Oxygen System to operate at its full potential.
Operating Temperature 50º F to 104º F (10º C to 40º C)
Operating Humidity 10% - 95% at an 82.4°F (28°C) dew point
Transport/Storage Temperature -4º F to 140º F (-20º C to 60º C)
Humidity: Up to 95% Non-Condensing
Electrical Use no extension cords or electrical
outlets controlled by a switch.
Altitude 0 – 13,123 feet (0 – 4000 meters)
Placement Do Not block air inlet or bottom exhaust vent
(keep at least 3 inches away from walls,
draperies, furniture, etc.)
Environment Smoke, pollutant, and fume free.
Operating Time Up to 24 hours per day when connected
to external power.
When moving the ECLIPSE from an extreme environment, allow time for your
ECLIPSE to acclimate to the recommend operating environment.
WARNING Operating your ECLIPSE outside the recommended operating
environment may impact performance, cause damage and will void
the warranty.
CAUTION Locate the unit in a well-ventilated room that provides adequate air
fl
ow and c
hec
k if air is being prevented from entering or exiting the
system by furniture, draperies, bedding, clothing or pets.
Recommended Operating Environments
Model 1000 User Manual
ECLIPSE
TM
Model 1000 User Manual
|
15