Symptom Probable Cause Remedy
ECLIPSE does not power on No Power Cartridge installed Install charged Power Cartridge
when ON/OFF button is pressed or plug into external power
Power Cartridge is discharged Plug into external power
Malfunction Contact your dealer/provider
No Oxygen Blocked or kinked tubing Check tubing or replace
Filter Blocked Clean Air Inlet Filter
Humidifier Check humidifier attachment
and tubing
ECLIPSE not ON Turn ECLIPSE ON
Tubing or cannula is not properly Check tubing, cannula and
connected or kinked connections
Power Cartridge is discharged Install charged Power Cartridge
or no external power is present or attach external power
Malfunction Contact your dealer/provider
Low Oxygen Concentration Restriction in tubing or humidifier Repair or replace tubing
or humidifier
Filter restricted Clean or replace air inlet filter
Malfunction Contact your dealer/provider
Low Oxygen Flow Restriction in humidifier or tubing Repair or replace humidifier
or tubing
Filter Blocked Clean or replace air inlet filter
Malfunction Contact your dealer/provider
No Oxygen delivered in Tubing/cannula longer than Attach 7 foot (2.1m)
Pulse Mode
7 feet (2.1m) tubing/cannula
Humidifier attached Remove humidifier
No inspiration detected Contact your dealer/provider
Power Cartridge Status Gauge Power Cartridge needs Fully discharge the Power
never indicates fully charged calibration Cartridge and recharge