EasyManuals Logo

ECS AdvancedIQ AG-2000 User Manual

ECS AdvancedIQ AG-2000
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
59
Handheld Gas Analyzer - Factors Influencing Accurate Readings
1. Elevation changes will affect the accuracy of the nitrogen purity readings. The deviation of the
nitrogen purity can be approximately 1% per 250 feet of elevation.
a. Calibration of the instrument should be performed when elevation at which the product
used changes more than 500 feet.
2. Temperature effects the accuracy of the nitrogen purity readings. The gas analyzer will hold
calibration and correctly read nitrogen purity ±3% when thermal equilibrium within the operating
temperature range. The device must be thermally stable when calibrated and allowed to
thermally stabilize after experiencing temperature changes before readings are accurate.
a. For best results, perform the calibration procedure at a temperature close to the
temperature where analysis will occur.
b. Allow adequate time for the sensor to equilibrate to a new ambient temperature.
CAUTION: “CAL Err St” may result from a sensor that has not reached thermal equilibrium.
Nitrogen Production Flow Rate Check Procedure
The production rate of the nitrogen generator can be verified using the production flow meter in the
nitrogen generator cabinet (See Generator Configuration Diagram).
1. With the nitrogen generator powered and in the nitrogen production mode:
a. Open the nitrogen production flow meter isolation ball valve to automatically start the
nitrogen generator.
b. Close the regulated and fast fill valves on the air maintenance device (AMD) to the fire
sprinkler system(s).
c. Read and document the nitrogen production flow rate on the nitrogen production flow meter.
i. Flow rate indicated is in standard cubic feet per hour (SCFH).
ii. The production rate is measured using the center of the silver ball in the flow meter.
d. Open the regulated valve on the air maintenance device (AMD) to the fire sprinkler system(s).
e. Close the flow meter isolation ball valve.
f. Compare the nitrogen production flow rate with the nitrogen production values listed in the
following chart.
Nitrogen Generator Flow Rates
Nitrogen Generator
Production Rate SCFH (L/min)
Bypass Mode
Nitrogen Generation Mode
AG-675 Wall Mount w/ Integral Air Compressor
150 (70.8)
20 (9.4)
AG-950 Wall Mount w/ Integral Air Compressor
198 (93.5)
27 (12.7)
AG-2000 Wall Mount w/ Integral Air Compressor
576 (271.8)
40 (18.7)
AG-3500 Wall Mount w/ Integral Air Compressor
576 (271.8)
90 (42.5)
AG-6500 Stand Alone w/ 7.5 hp Air Compressor
1,458 (681.1)
240 (113.3)
AG-11000 Stand Alone w/ 7.5 hp Air Compressor
1,458 (681.1)
240 (113.3)
AG-18500 Stand Alone w/ 7.5 hp Air Compressor
1,458 (681.1)
425 (200.6)
AG-22500 Stand Alone w/ 10 hp Air Compressor
2,100 (991.1)
550 (259.6)
NOTE: If production flow rate is lower than flow rates in the production chart, check pressure regulator
of the excess bleed off assembly.
If production flow rate identified and documented varies more than 10% of the flow rates in the
production chart, Contact ECS.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ECS AdvancedIQ AG-2000 and is the answer not in the manual?

ECS AdvancedIQ AG-2000 Specifications

General IconGeneral
BrandECS
ModelAdvancedIQ AG-2000
CategoryPortable Generator
LanguageEnglish

Related product manuals