-----------Installation
Procedures------------
R.OW
CONTROL
PRESSURE
REGULATOR
(IF
REaD)
Fl.t:IN
CONTROL
BYPASS
R.OW (IF
REO'D}
INCREASED TOTAL
SMoFLE
R.ON FOR
FASTER
RESPONSE
EXHAUST
Figure2-1.
Suggested
sampliDg
system
.
the
installation,
and the user must ensure
that
no
portion
of the sample line is at a temperature lower thanthe
highest dew point
anticipate
d
If
electrical
heater
lines
are
used
,
it is usually
sufficient
to connect themto
a
variable
transformer
to adjust
the
heating
level
If
the sample lines
are long,
it
may be desirable to
wrap
them
in
insulation
to minimize the
amo1m.t
ofheat required for
preheating
.
The
Senso
r
temperature must always
be
maintained
at least 5°C above the dew point temperature
of the gas
sample
.
The
temperature
ofthe
Sensor
must
not
exceed+
1
00°C.Heating
the Sensor and sample
lines
above the dew point of the
gas
sample does not
change
the dew point of the
sample
.
2.10 SELECTION OF COMPONENTS
AT
LOW DEW
POINTS
Whenmeasuring dew-points below
-l8°C,
increased
attention to sampling
details
must
be
made
as
th
e
dew
point
is
lowered.
Rubber,neoprene,
tygon, andmost plastic tubing
are
lm.desirable
for use
in
hygrometer
system
s
as sample
lines from the
sampling
system to the
Sensor.
Specifi-
cally,
nonhygroscopic
tubing,such as
stainless
steel
,
Teflon, or
KEL-F
,
with a minimum of
joints,
fittings,
and other plumbing, is
recommended.
The
actual
selection of the sample line material should be based
on
2-4