Instrument
is
completely
inoperative
with no LEDs
lit.
Check the 120/240 voh PC card for proper position for
input
AC
voltage
being used
If
the
fuse is good, check for line voltage across the
AC
terminals.
Line voltage and
fuses
are good, but
still
no
indication of
equipment
operation.
Unplug the power cord and slide the clear plastic door to the left to check
the
AC voltage select
card
and fuse. Check
that
the cables connecting to
the
PWB
are
intact.
Replaced fuse blows
when power is
turned
on
again.
Check that the proper fuse is being used. The fuse value should have
a
2-amp rating for 120 VAC
operation
and
a l-amp
rating for 240
VAC
operation.
Both
fuses
should be rated 250V,
3
AG,
Slo-Blo.
CONTROL
indicator
is
too far left or too
far
right daring the
time
the ABC indicator
is
on
steady.
This
indication
can occur
if
the
Sensor is not plugged
in,
if
the
Sensor
has
been
changed,
or
if
the Sensor mirror
sur:W:e
is very dirty.A dirty
mirror
causes the CONTROL indicator to indicate more to the left. Adjustment
of
BALANCE
potentiometer RS2 can usually bring this indication
back.
to
center scale
if
a new Sensor
has
been
attached.
Cleaning a dirty Sensor
mirror
has
a
similar
effect.
Control indicator
and
dew point
temperature
output do not settle
out
to a steady
reading.
First check that the dew poiut sample flowing through the Sensor is under
a
constant
pressure
and does not vuy,
and
that
there
are no plmnbing
leaks.
If
thes
e
inputs to the
Seusor
check
out, then either the GAIN
or
COMPENSATION
control may be set too high, or the sample flow rate
may
be too
high.
Sluggish response
to
changes in dew
point.
TIDCKNESS
setting may be too high or
GAIN
setting may be too
low
.
Dew point reading
is
always too
high.
THICKNESS
setting
is too low or GAIN settingis too
low.
When operating
at
high Sensor
ambient
temperatures
near
+IOOOC
,
the
instrument
indicates
ambient
temperature
rather
than
dew
point
temperature.
If
the
Sensor base temperature
has
been
allowed to
exceed+
1
OSOC,
the
thermostat
in the Sensor will open, preventing the mirror from cooling to
the
dew
point
The mirror temperature will now rise to ambient. Reduce
the
Sensor temperature to +800C or below to reset the Sensor
thermostat
Model2000 Series DewPrime
Dew
Point
Hygrometers
----------Troubleshooting
and
Repair---------
Table
6-1. Troubleshootin
g
clw1:
preliminary
ehecks.
.
6-2