4. To measure the set-point pressure accurately:
Connect pin 7 to pin 2 (or to ground)
Use a voltmeter to measure the voltage between
pins 3 and 5
Use the data in Tables 2 and 3 to convert the
voltage to an equivalent pressure
If necessary, turn the set-point potentiometer to
change the set-point pressure
Disconnect pin 7 from pin 2 (or from ground)
4. Genaues Messen des Druck-Schaltpunkts:
Pin 7 an Pin 2 (oder Masse) anschließen.
Mit einem Voltmeter die Spannung zwischen
Pin 3 und 5 messen.
Mit Hilfe von Tab e l l en 2 und 3 die Spannung in
den entsprechenden Druck umrechnen.
Nötigenfalls das Schaltpunkt-Potentiometer
drehen, um den Druck-Schaltpunkt zu ändern.
Pin 7 von Pin 2 (oder Masse) trennen.
4. Pour mesurer la pression du point de consigne avec
précision :
Raccordez la broche 7 à la broche 2 (ou à
la terre).
Utilisez un voltmètre pour mesurer la tension
entre les broches 3 et 5.
Servez-vous des données des Tableaux 2 et 3 pour
convertir la tension en son équivalent en
pression.
Si nécessaire, tournez le potentiomètre du point
de consigne afin de modifier la pression du point
de consigne.
Déconnectez la broche 7 de la broche 2 (ou de
la terre).