EasyManua.ls Logo

Edwards EH1200

Edwards EH1200
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
M518-00-880 Issue A
Page 12 © Edwards Limited 2012. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
Minimum supply flow rate for
T160/T4 operation
10.5
(max 20°C water: CXS160/250
+ EH1200)
12.5
(max 20°C water:CXS250
+ EH2600)
13.5
(max 25°C water: CXS160/250
+ EH1200)
16.5
(max 25°C water:CXS250
+ EH2600)
l/min
Supply temperature range for T3
operation
5-35 °C
Supply temperature range for
T160/T4 operation
5-20, 5-25 - see min required
pressure differential
°C
Water type Treated or non-corrosive industrial
EH1200 Water flow switch setting 2 l/min
EH2600 Water flow switch setting 4 l/min
Maximum particle size 0.03 mm
2
Acidity 6.5 to 8.0 pH
Hardness <100 ppm of CaCO
3
(<100 mg of CaCO
3
per
litre)
Resistivity 1k ρ 1000k ohm-cm
Materials in contact with
cooling-water
Stainless steel, Nitrile, PTFE, brass
,
polyamide and fluoroelastomer
Water inlet connection 3/8 inch BSPT male
Water outlet connection 3/8 inch BSPT male
*
The TMS is a valved system and the “minimum required pressure differential” may not correlate with water
flow rates under all operating conditions. The “minimum required pressure differential” is required to
maintain adequate cooling-water flow under adverse operating conditions.
Based on water with a heat capacity of 4.18 kJ/kgK. Flow compensation must be made for coolants with a
lower value.
Table 7 - Water cooling system data (continued)
Description Rating Units