M518-00-880 Issue A
Page 12 © Edwards Limited 2012. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Technical data
Minimum supply flow rate for
T160/T4 operation
10.5
†
(max 20°C water: CXS160/250
+ EH1200)
12.5
†
(max 20°C water:CXS250
+ EH2600)
13.5
†
(max 25°C water: CXS160/250
+ EH1200)
16.5
†
(max 25°C water:CXS250
+ EH2600)
l/min
Supply temperature range for T3
operation
5-35 °C
Supply temperature range for
T160/T4 operation
5-20, 5-25 - see min required
pressure differential
°C
Water type Treated or non-corrosive industrial
EH1200 Water flow switch setting 2 l/min
EH2600 Water flow switch setting 4 l/min
Maximum particle size 0.03 mm
2
Acidity 6.5 to 8.0 pH
Hardness <100 ppm of CaCO
3
(<100 mg of CaCO
3
per
litre)
Resistivity 1k ≤ ρ ≤ 1000k ohm-cm
Materials in contact with
cooling-water
Stainless steel, Nitrile, PTFE, brass
,
polyamide and fluoroelastomer
Water inlet connection 3/8 inch BSPT male
Water outlet connection 3/8 inch BSPT male
*
The TMS is a valved system and the “minimum required pressure differential” may not correlate with water
flow rates under all operating conditions. The “minimum required pressure differential” is required to
maintain adequate cooling-water flow under adverse operating conditions.
†
Based on water with a heat capacity of 4.18 kJ/kgK. Flow compensation must be made for coolants with a
lower value.
Table 7 - Water cooling system data (continued)
Description Rating Units