User Manual EickSonic II ENGL
22/28
Fault
Possible causes
Solutions
The vibration of the tip
becomes weak.
The tip is loose by because of
vibration.
Screw on the tip tightly
(as shown in picture 4).
The coupling between the
handpiece and the cable is
not dry.
Dry it by the heated wind.
The tip is damaged [Note 3].
Replace with a new one.
The vibration is too strong and
the vibrating intensity control
knob is malfunction.
The vibrating intensity control
knob is damaged.
Contact EICKEMEYER
®
There is water seeping from
the coupling between the
handpiece and the cable.
The waterproof "O" ring was
damaged.
Replace with a new "O" ring.
If the problem still cannot be solved yet, please contact EICKEMEYER
®
.
[Note 1] The water control knob can adjust the water volume according to the symbol.
[Note 2] To clean the water pipe with the multi-function syringe of the dental unit (as showed in the
picture 5):
Picture 5
-
Snip the water pipe with scissors at a distance of 1015 cm from the water entrance.
-
Turn on the power switch, get through to the power.
-
Connect the multi-function syringe of dental unit to the water pipe.
-
Step on the foot switch.
-
Turn on the switch of the multi-function syringe, press the air or water into the water line in the
machine, then the impurity blocked in the water line can be eliminated.
[Note 3] If the scaling tip has been screwed on tightly and there is fine spray too, the following
phenomena show that the scaling tip was damaged:
The vibrating intensity and the water atomization degree become weak obviously. When operating, there
is some buzz when the scaling tip is working.