3
1. Unplug the device from the electrical outlet.
2. Press the ON/OFF switch to the ON position. The LCD should display the
main menu.
ENT
ENTER FUNCTION:
READ 1.1 (OR 0.5) SPIN
RUN
3. Press the RUN and ENT buttons simultaneously and hold them down until
the battery pack capacity is displayed.
When there is 20% capacity remaining, the main menu LCD will flash as
an indication that the battery pack is beginning to run down. It will still be
possible to operate the device for a few spins but the operator should
recharge the battery pack soon. Once there is no longer sufficient
capacity to operate the instrument, it will fail to complete a cycle and the
message CHARGE BATTERY will be displayed.
4. If the device is plugged into an electrical outlet with the ON/OFF switch in
the ON position and the RUN and ENT buttons are pressed
simultaneously, the LCD will display either CHARGING when in the
process of charging or TRICKLE CHARGING if fully charged.
REGARDLESS OF WHAT THE LCD MAY DISPLAY AT ANY GIVEN
TIME DURING THE CHARGING CYCLE, IT IS IMPORTANT TO
ALWAYS ALLOW THE BATTERY PACK TO CHARGE OVERNIGHT OR
AN EQUIVALENT PERIOD OF TIME TO ENSURE A FULL CHARGE.
A FULLY CHARGED BATTERY PACK WILL GENERALLY PROVIDE 75
SPIN CYCLES IF THE CENTRIFUGE IS TURNED OFF BETWEEN
TESTS.
BATT. CAPACITY
XX %