SECTION 3
SETUP PROCEDURE
3.1 DISPLAY OPTIONS
This setup procedure will allow the operator to store several functions in the display
option memory until a change is required. These functions include the selection of:
• Tube size - 1.1mm or 0.5mm Inside Diameter (ID). The device must be set to
the tube size being used. (The ID of capillary tubes may vary slightly - up to
0.1mm.)
• Display language - English, Spanish, French, German or Italian
• HCT value suffix - Choice of displaying % sign or only the value itself
• Decimal point - Choice of displaying hematocrit in whole numbers or with a
decimal point.
Performing this setup procedure eliminates the need to enter these settings into the
device each time a test is performed.
3.2 SETUP MENU AND PROCEDURE
A. With the power ON, the LCD will display the main menu:
ENT
ENTER FUNCTION:
READ 1.1 (OR 0.5) SPIN
RUN
B. Simultaneously press and release the ENT and RUN buttons. The LCD
will change to:
ENT
SETUP
TUBE I.D. NEXT
RUN