EKO INSTRUMENTS CO., LTD. - Pyranometers MS-802/802F/402/402F/602 - Instruction Manual Ver.13
1) Use the relative measurement uncertainty of the calibration factor to determine the measurement
error of each single measurement.
2) Make averages of a large number of individual measurements (for example 1-second
measurements averaged over 5 minutes).
3) Adopt reported best-practice measurement uncertainties (see for example [Diffuse IOP],
[Pyrheliometer IOP].
Traceability
Internal reference pyranometer maintained at EKO is traceable to the absolute cavity which is directly
compared against the WRR (World Radiometric Reference) Primary Standard (Absolute Cavity) managed at
PMOD (Davos, Switzerland). The logger system used for the calibration measurement is traceable to
JEMIC (Japan Electric Meters Inspection Cooperation).
Internal reference pyranometer is calibrated directly compared against the pyrheliometer, which is measured
against the EKO absolute cavity, and 2 units of internal reference pyranometers by Shading Method ([A New
Method for Calibrating Reference and Field Pyranometers (1995)] Bruce W Forgan) every one year.
EKO absolute cavity is directly traceable to WRR by comparing against WRR every 5 years.