Page 8 of 24 MEL-TEMP M8160 Issue 1.5
5. INSTALLATION.
5.1. Electrical Safety and Installation.
5.1.1. This equipment is designed to safe under the following conditions:-
Indoor use.
Altitude up to 2000 meters.
Temperatures between 5 C and 40 C.
Maximum relative humidity 80% for temperatures up to 31 C decreasing linearly to
50% relative humidity at 40 C.
Mains supply voltage fluctuations up to 10% of the nominal voltage.
Transient overvoltage’s typically present on the mains supply.
Applicable rated pollution degree 2.
5.1.2. Check the voltage on the PSU data label. Ensure the rating conforms to your local
supply.