PURELAB Option-R 7/15
Operator Manual

Questions and Answers

  • T
    Timothy WagnerAug 4, 2025
    How to fix no display message on ELGA PURELAB Option-R15?
    • B
      Bethany PerryAug 4, 2025
      If there is no display message on your ELGA Water Dispenser, verify the mains supply and lead are properly connected. Ensure the mains power is switched on. Also, inspect the fuse in the power inlet module and replace it if it's blown.
  • R
    romancrystalAug 8, 2025
    What to do if ELGA PURELAB Option-R15 output flow is outside specification?
    • R
      Ryan SimsAug 8, 2025
      If the output flow of your ELGA Water Dispenser is outside the specified range, start by checking the supply pressure. Then, compare the RO flow-rate of your unit against the graphs detailed in Section 3.3 of the manual, paying attention to treated water output in relation to temperature and feedwater pressure. Ensure the inlet strainer and receiver strainer are clean. If necessary, consider contacting a service technician to install or replace an optional booster pump.
  • K
    kylebellAug 12, 2025
    What to do if the reservoir low level audible alarm sounds on my ELGA PURELAB Option-R15?
    • C
      Connie CoxAug 12, 2025
      If the reservoir low level audible alarm sounds on your ELGA Water Dispenser, you can press the crossed bell button to mute the alarm. The reservoir should then automatically refill. Check that the process mimic shows the reservoir filling. Also, verify the feedwater supply and the connections to the reservoir.
  • M
    Mary GarciaAug 16, 2025
    How to fix no flow from dispense tap on ELGA Water Dispenser?
    • T
      tsmithAug 16, 2025
      If there is no flow from the dispense tap of your ELGA Water Dispenser, push the PROCESS button. Also, ensure that the reservoir level is above 40%.
  • A
    Anthony HaleAug 21, 2025
    What to do if UV lamp failure audible alarm sounds on ELGA PURELAB Option-R15?
    • F
      fernandezstevenAug 21, 2025
      If the UV lamp failure audible alarm sounds on your ELGA Water Dispenser, you can press the crossed bell button to mute the alarm. Check that all electrical connections are secure. When applicable, follow the UV lamp replacement procedure. You can also temporarily continue operation without the UV lamp if necessary.
  • R
    Randy WolfeAug 25, 2025
    How to stop ion-exchange cartridge replacement alarm on ELGA PURELAB Option-R15 Water Dispenser?
    • E
      Emily WoodardAug 26, 2025
      If the ion-exchange cartridge replacement alarm is active on your ELGA Water Dispenser, replace the ion-exchange cartridge pack.
  • E
    Edward LawrenceAug 29, 2025
    What to do if pre-treatment cartridge replacement alarm sounds on ELGA Water Dispenser?
    • M
      Matthew TurnerAug 29, 2025
      If the pre-treatment cartridge replacement alarm is active on your ELGA Water Dispenser, replace the pre-treatment cartridge.

Summary

INTRODUCTION

Product Range

Lists the different PURELAB Option-R models available.

Use of this Manual

Provides guidance on how to use the operator manual effectively.

Customer Support

Details how to obtain service support and consumable items.

HEALTH AND SAFETY NOTES

General Safety Precautions

Highlights critical safety information regarding electricity, pressure, UV light, and chemicals.

PRODUCT AND PROCESS DESCRIPTION

Product Overview

Describes the PURELAB Option-R unit and its purpose.

Process Description

Explains the water purification stages and technologies used by the unit.

Technical Specifications

Details electrical requirements, dimensions, connections, and performance data.

CONTROLS

Control Panel Overview

Explains the function of the control panel, buttons, and display icons.

INSTALLATION INSTRUCTIONS

Unpacking the PURELAB Option-R

Covers the initial steps of unpacking the unit.

Positioning the PURELAB Option-R

Provides guidance on selecting a suitable location and environment for the unit.

Connecting up the PURELAB Option-R

Provides instructions for connecting water tubes, electrical supply, and reservoir level control.

Initial Controller Set-Up

Guides through setting up system preferences like auto-restart, alarms, and water purity units.

Initial Start Up

Details the procedure for the first-time start-up, including system rinsing.

OPERATION

Dispensing and Intermittent Mode

Explains how to dispense water and the automatic intermittent mode for maintaining purity.

Alarm Conditions Overview

Describes various alarms, their causes, and how to acknowledge them.

MAINTENANCE

Identification of Consumables

Identifies the unique replacement consumables designed for the unit.

Frequency of Consumable Replacement

Recommends the service life for consumables based on typical usage.

Replacing the LC140 Pre-treatment Cartridge

Step-by-step guide for replacing the LC140 pre-treatment cartridge.

Replacing the LC141 lon-exchange Cartridge Pack

Instructions for removing and replacing the LC141 ion-exchange cartridge pack.

Replacing the Ultraviolet Lamp (LC105)

Procedure for changing the LC105 ultraviolet lamp.

Cleaning the Inlet Strainer

Details on how to check and clean the inlet strainer.

Cleaning the Re-Circulation Strainer

Details on how to check and clean the re-circulation strainer.

Replacement of LC143 Reverse Osmosis Cartridge(s)

Information on when to replace the LC143 reverse osmosis cartridge.

SANITIZATION PROCEDURES

Sanitization Process Steps

Outlines the complete procedure for sanitizing the unit and reservoir.

Sanitization of Unit and Reservoir

Specific steps for sanitizing both the unit and the reservoir.

CT1 Sanitization Tablet Safety Information

Provides safety information and handling precautions for the CT1 sanitization tablet.

TROUBLE SHOOTING

Common Problems and Solutions

Lists common issues, their causes, and recommended actions for resolution.

CONSUMABLES AND ACCESSORIES

Product Listings

Catalogues of available consumables, their service life, and accessories with part numbers.

KEY TO CONTROL PANEL

Icons and Their Meanings

Defines the various icons used on the control panel.

Alarm Conditions

Details the icons and conditions that signify specific alarms.

Replacement Timers

Explains the icons and presets related to cartridge replacement timers.

Low Level, Quality and Standby Alarms

Shows screen examples for low level, quality, and standby alarms.

WARRANTY;CONDITIONS OF SALE

General Limited Warranty Details

Outlines the warranty terms for manufactured products.

Water Systems Limited Warranty Details

Specifies the warranty terms for water systems, excluding membranes and cartridges.

USEFUL CONTACT DETAILS

Contact Information and Resources

Provides contact details for ELGA LabWater and additional resources.

ELGA PURELAB Option-R15 Specifications