EasyManua.ls Logo

Elitech DMG-4B - Page 26

Elitech DMG-4B
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. Set the refrigerant, (long press + unlock, then short press
or switch the refrigerant type).
3.3 Pressure Holding Measurement Mode(HOLD)
Note: This function needs to be purchased before it can be used, and
the purchase method is detailed in 4.4 of this article.
The high-voltage side channel is the system pressure-holding test
terminal.
1.Connect the high-pressure side refrigerant hose to the system
under test, and close the high-pressure valve.
2.Enter the holding pressure test : press the and button at
the same time , the upper left corner of the window displays HOLD
(Figure 2).
3.Set the pressure holding parameters :
Pressure alarm value: When the current pressure value reaches the
target pressure holding value, the device will alarm, and the
pressure holding can be started at this time.
Decay ratio: adjust the decay ratio, and set the pre-warning value for
pressure keeping.
Temperature Compensation: Turn on/off the temperature
compensation switch.
Pressure hold time: set the hold time.
Figure 2 Enter the holding pressure test interface
!
The parameters of this step need to be connected to the APP for
setting
!

Related product manuals