6.5.7 Cleaning Items (Treating Samples) in Auto Mode
NOTICE
To stop ultrasonics at any time, press the Sonics key.
Table 6.21 Treating Samples in Auto Mode
Step Action
1 Turn Main Power switch on.
2 Press the On/Standby key to turn on the unit.
3 If necessary, degas the liquid. See 6.5.3 Degassing for instructions.
4
Set the amount of time you wish the items to be cleaned:
Use Up/Down keys to increase/decrease cycle time (hold for quick
increments/decrements)
5
Set the tank temperature:
Press the Fn key until the Set Temperature icon flashes
Press the Up/Down keys to alter the setting to the tank temperature you wish
to reach
Press the Heat key once to activate heat. The Heat icon appears
Units can be switched between °F or °C using the Fn key. See Fn key
description on
6.5.1 CPXH Series Explanation of Controls
6
Set the ultrasonic power level:
Press the Fn key until the Power Level icon flashes
Press the Up key to select high power ultrasonics or press the Down key to
select low power ultrasonics
7
Press the Auto key to begin Auto Cycle. Heater will turn on and sonics will
start once set temperature is attained.
8
Place the items into a basket, perforated tray, or beakers in a positioning
cover.
9
If using beakers or a solid tray, add cleaning solution to beakers or tray
to cover the items.
10
Slowly lower the tray or beakers into the tank. Do not allow items to
contact the tank bottom.
11 When items are clean, slowly remove them from the tank.
12 Rinse clean items with clean, warm water and dry, if necessary.

Table of Contents