Product Data Sheet
00813-0100-4530, Rev DA
December 2010
Rosemount 5300 Series
24
Cryogenic Temperature tank connection (Operating Temperature and Pressure code C):
ASME / ANSI Flange Rating 316L SST flanges according to ASME B16.5 Table 2-2.3:
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
HP/C: Class 2500 up to max 200 °C
HTHP: Class 2500 up to max 400 °C
Alloy C-276 (UNS N10276) flanges according to ASME B16.5 Table 2-3.8:
HP/C: Class 1500 up to max 200 °C
HTHP: Class 1500 up to max 400 °C
EN Flange Rating 1.4404 according to EN 1092-1 material group 13E0:
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
HP/C: PN 320 up to max 200 °C
HTHP: PN 320 up to max 400 °C
Alloy C-276 (UNS N10276) flanges according to EN 1092-1 material group 12E0
Fisher & Masoneilan Flange Rating 316L SST according to ASME B16.5 Table 2-2.3:
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
HP/C: Class 600 up to max 200 °C
HTHP: Class 600 up to max 400 °C
JIS Flange Rating 316L SST according to JIS B2220 material group 2.3:
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
HP/C: Max temp. 200 °C. Final rating depends on flange.
HTHP: Max temp. 400 °C. Final rating depends on flange.
Tri-Clamp Rating Maximum pressure is 16 bar for 1.5 in. (37.5 mm) and 2 in. (50 mm) housing; and 10 bar for 3 in. (75
mm) and 4 in. (100 mm) housing. The final rating depends on the clamp and gasket. Tri-Clamp is
available for the Standard Temperature and Pressure seal.
Plate Design Certain models of flanged Alloy and PTFE covered probes have a tank connection design with a
protective flange plate of the same material as the probe and with a backing flange in 316L / EN
1.4404. The protective flange plate prevents the backing flange from being exposed to the tank
atmosphere.
Flange rating according to SST backing flange ASME B16.5 Table 2-2.3, EN 1092-1 material group
13E0, and JIS B2220 material group 2.3.
Alloy C-276 protective plate:
Standard: Max. 302 °F/580 psig (150 °C/40 Bar). Flange plate design is available up to Class
300/PN 40
HP: Max temp. 200 °C. Flange plate design is available up to Class 600/PN 63
HTHP: Max temp. 400 °C. Flange plate design is available up to Class 600/PN 63
Alloy 400 protective plate:
Standard: Max. 302 °F/580 psig (150 °C/40 Bar). Flange plate design is available up to Class
300/PN 40
PTFE protective plate:
Standard: Max. 302 °F/580 psig (150 °C/40 Bar)
Flange Connection Rating See Table 5 and Table 6 for the conditions used for flange strength calculations
-200 (-129)
-320 (-196)
Max. Rating, C Connections
Pressure
psig (bar)
Temp.
°F (°C)

Other manuals for Emerson Rosemount 5300 Series