Product Data Sheet
00813-0100-4530, Rev DA
December 2010
Rosemount 5300 Series
32
Material Exposed To
Tank Atmosphere
Standard Probe (Operating Temperature and Pressure code S):
• Material model code 1: 316L SST (EN 1.4404), PTFE, PFA, and O-ring materials
• Material model code 2: Alloy C-276 (UNS N10276), PTFE, PFA, and O-ring materials
• Material model code 3: Alloy 400 (UNS N04400), PTFE, PFA, and O-ring materials
• Material model code 7: PTFE
• Material model code 8: PTFE, 316 L SST (EN 1.4404), and O-ring materials
HTHP Probe (Operating Temperature and Pressure code H):
• Material model code 1: 316L SST (EN 1.4404), Ceramic (Al2O3), Graphite, and Inconel
• Material model code 2 and H: Alloy C-276 (UNS N10276), Ceramic (Al2O3), Graphite, Inconel
HP Probe (Operating Temperature and Pressure code P):
• Material model code 1: 316L SST (EN 1.4404), Ceramic (Al2O3), Graphite, PFA, PTFE, Inconel
• Material model codes 2, H: Alloy C-276 (UNS N10276), Ceramic (Al2O3), Graphite, PFA, PTFE,
Inconel
Cryogenic Probe (Operating Temperature and Pressure code C):
• Material model code 1: 316L SST (EN 1.4404), Ceramic (Al2O3), Graphite, PFA, PTFE, Inconel
Pressure Equipment Directive (PED)
Complies with 97/23/EC article 3.3
Total Probe Length This is defined from the upper reference point to the end of the probe (weight included, if applicable).
Select the probe length according to the required measuring range (the probe must be hung and fully
extended through the entire distance where level readings are desired).
Most of the probes can be cut in field. However, there are some restrictions for the standard and
HP/C coaxial probes: these can be cut up to 2 ft. (0.6 m). Probes shorter than 4.1 ft. (1.25 m) can be
cut to the minimum length of 1.3 ft. (0.4 m). The HTHP coaxial probe and the PTFE covered probes
cannot be cut in the field.
Minimum and Maximum Probe
Length
Coaxial: 1.3 ft (0.4 m) to 19.7 ft (6 m).
Rigid Twin Lead: 1.3 ft (0.4 m) to 9.8 ft (3 m).
Flexible Twin Lead: 3.3 ft (1 m) to 164 ft (50 m).
Rigid Single Lead (0.3 in./8 mm): 1.3 ft (0.4 m) to 9.8 ft (3 m)
Rigid Single Lead (0.5 in./13 mm): 1.3 ft (0.4 m) to 14.8 ft (4.5 m)
Flexible Single Lead: 3.3 ft (1 m) to 164 ft (50 m)
Probe Angle 0 to 90 degrees from vertical axis
Tensile Strength 0.16 in. (4 mm) Flexible Single Lead probe: 2698 lb (12 kN)
0.24 in. (6 mm) Flexible Single Lead probe: 6519 lb (29 kN)
Flexible Twin Lead probe: 2023 lb (9 kN)
Collapse Load 0.16 in. (4 mm) Flexible Single Lead probe: 3597 lb (16 kN)
0.24 in. (6 mm) Flexible Single Lead probe: 7868 lb (35 kN)
Sideway Capacity Coaxial probe: 73.7 ft. lbf, 3.7 lb at 19.7 ft. (100 Nm, 1.67 kg at 6 m)
Rigid Twin Lead: 2.2 ft. lbf, 0.22 lb at 9.8 ft. (3 Nm, 0.1 kg at 3 m)
Rigid Single Lead: 4.4 ft. lbf, 0.44 lb at 9.8 ft. (6 Nm, 0.2 kg at 3 m)
Maximum Recommended Nozzle
Height
4 in. (10 cm) + nozzle diameter
For coaxial probes, there are no restrictions
Total
Probe
Length
NPT
BSP/G
Flange
Tri-Clamp
Upper
Reference
Point