Captair Smart
Replacing the HEPA H14 / ULPA lters
Pre-requisites
The operative responsible for replacing the lter is kept up-to-date with the exhaustive list of products handled in
the fume hood by the user so that the correct EPI can be used
The laboratory is empty when the operation is carried out
The laboratory is ventilated by mechanical or natural means while the operation is carried out
Minimum protective equipment
One-piece overall + overshoes + bouffant cap
Laboratory gloves (latex or nitrile)
Protective glasses
Breathing mask with particle lter (P3)
Notice: additional equipment could be required.
This procedure is applicable to HEPA/ULPA lters located at the bottom of the ltration columns and designed to trap
powders handled inside the device enclosure.
Strict chronological order to follow:
1- Switch on the device fan
2- Carefully spray the bottom surface of the HEPA/ULPA lter
(paint with NON FLAMMABLE propellent), to be done inside the enclosure
3- Allow at least 5 minutes with the fan running for the spray to dry
4- Shut down and unplug the device and disconnect the fan module power supply cable and the sampling tubes from the
sampling area (if installed)
5- Carefully remove the molecular lter(s) (if present) and the fan module
6- Carefully unwrap the new HEPA/ULPA lter
Keep the plastic lm and cardboard box so that you can use it later to pack up the used lter
Lay out the lm on a at surface in the immediate vicinity of the operation so that it is at the ready
7- Carefully remove the used HEPA/ULPA lter and immediately place it contaminated-side down onto the plastic lm
8- Clean the lter housing and the enclosure (using water + surfactant)
9- Package up the used lter + contaminated equipment
Seal the plastic lm tightly
10- Place the sealed package in the box the new carbon lter came in, then seal it using adhesive tape, clearly write «used
lter» on the packaging.
Have the lter disposed of via a suitable disposal process in accordance with the applicable regulations.
To nd out more, please contact your usual advisor.
11- Fit the new HEPA/ULPA lter, main molecular lter (if present) and the fan module, followed by the backup molecular
lter (if present). Make sure that all the column components
12- Reconnect the device’s various cables and hoses, switc h the device back on and check the air speed using the
anemometer calibration procedure via the administrator interface
Type AS
For organic vapors
Type BE +
Multi-application for acid and organic vapors
Type K
For ammonia vapors
Type F
For formaldehyde vapors
HEPA H14
For powders
ULPA
For powders
The table below summarises the different types of carbon lters that Erlab® offers along with their elds of application.