#550 DIGITAL INSULATION TESTER
INSTRUCTION MANUAL
SAFETY INFORMATION
Read the following safety information carefully before attempting to operate or service the meter.
To avoid damaging the instrument, do not apply signals which exceed the maximum limits shown in the
technical specications tables.
Do not use the meter or test leads if they look damaged. Use extreme caution when working around bare
conductors or bus bars.
Accidental contact with the conductor could result in electric shock.
Use the meter only as specied in this manual; otherwise, the protection provided by the meter may be
impaired.
Read the operating instructions before use and follow all safety Information.
Caution when working with voltages above 60V DC or 30V AC RMS. Such voltages pose a shock hazard.
Before taking resistance measurements or testing acoustic continuity, disconnect circuit from main power
supply and all loads from the circuit.
Safety symbols:
Caution refer to this manual before using the meter.
Dangerous voltages.
Meter is protected throughout by double insulation or reinforced insulation.
When servicing, use only specied replacement parts.
CE
Comply with EN-61010-1
1. SPECIFICATIONS
1-1 General Information
Environment conditions: Installation Categories II
Pollution
Degree 2
Altitude up to 2000 meters
Indoor use only
Relatively humidity 80% max.
Operation Ambient 0~40ºC
Maintenance & Clearing:
Repairs or servicing not covered in this
manual should only be performed by qualied personnel.
Periodically wipe the case with a dry cloth. Do not use abrasives or solvents on this
instrument.
Display: Large LCD with dual display
Measurement Range: 200
Ω
, 200k
Ω
, 200M
Ω
/250V, 200M
Ω
/500V,
2000M Ω /1000V, 750V/ACV, 1000V/DCV.
Sampling Rate: 2.5 times per second.
Zero Adjustment: Automatic adjust
ment.
Over Range Indicator: Number 1 of highest digit is displayed.
Low Battery Indication: The “ “ symbol is displayed when the battery
voltage becomes too low.
Operating Temperature: C to 4C (32ºF to 104ºF)
Question and Answer IconNeed help?

Do you have a question about the ES 550 and is the answer not in the manual?

ES 550 Specifications

General IconGeneral
BrandES
Model550
CategoryTest Equipment
LanguageEnglish

Summary

Safety Information

Specifications Overview

General Information

Specifies environmental conditions such as installation categories, pollution degree, altitude, usage, humidity, and operating ambient temperature.

Maintenance and Cleaning

Details procedures for repairs, servicing, and cleaning the instrument, emphasizing qualified personnel and specific cleaning methods.

Electrical Specifications

Provides detailed accuracies, ranges, circuit voltage, and overload protection for various measurement types like OHMS, voltage, and insulation resistance.

Parts and Controls

Battery and Test Lead Procedures

Connecting Test Leads

Instructions for correctly connecting test leads to the meter for MΩ range and Ω/ACV ranges.

Battery Check and Replacement

Guidance on checking battery status, identifying the low battery symbol, and replacing batteries with the correct type.

Test Lead Functionality Check

Procedure to verify test lead condition by checking resistance readings on the 200 Ω range.

Measurement Procedures

Insulation Resistance Measurements (200MΩ/250V)

Details performing insulation resistance tests at 200MΩ/250V, including minimum acceptable resistance values.

Insulation Resistance Measurements (2000MΩ/1000V)

Explains insulation resistance testing at 2000MΩ/1000V, voltage selection, and potential component damage concerns.

Lock Power On Feature

Describes the integrated lock feature for continuous operation of the test button without continuous pressing.

Low Resistance and Continuity Measurements

Instructions for measuring low resistance and continuity, including beep indication for low resistance values.

AC/DC Voltage Measurements

Guidance on performing AC and DC voltage measurements by connecting the meter in parallel to the circuit.

Power Tools and Small Appliances Testing

Instructions for testing power tools and small appliances, including specific advice for motor testing.

Motor and Cable Testing

Motor Testing (AC/DC)

Procedures for testing AC and DC motors, including connections to housing, terminals, and individual components.

Cable Insulation Testing

Instructions for checking cable insulation resistance to ground, lead sheath, and between conductors in pairs.