15
1.4 Replacement of Filter
1.4.1 SCR-A
1.
Unscrew and remove front cover
2.
Unscrew and remove the perforated shelf
3.
Unscrew and remove the UV lamp and unsnap the holder
4.
Unbolt and remove the filter clamp. Makesure to hold the filter when doing this.
5.
Carefully remove the filter and replace
6.
Reverse above steps
7.
Re-certify the cabinet