Calibration mV - AG-Select B1
From main screen select OK and press PLUS button
X 3 or until “CAL” is shown
Press OK button to go to 1
st
calibration point 650mV
Place clean mV electrode into the buffer solution,mix slowly and
then wait until display stabilises after around 1 minute. Use the
PLUS and MINUS buttons to trim the reading to mV 650
Press OK button to confirm and to go to next calibration point for
“ZERO” mV. To simulate 0 mV use either a shorting plug or using
a small piece of wire connect the outer body of the BNC
connector to the inner core section.
Proceed as for CAL1 and press OK button when done and press
ESC button to navigate back to start up screen