INSTRUMENT OPERATION - Simply aim the RayTemp 8 Infrared (IR) Thermometer at the target
and press the trigger to measure the surface temperature. The scan icon ' ' will be displayed.
When the trigger is released the reading is automatically held for 60 seconds and 'HOLD' will be
displayed, after which the thermometer will shut down.
MEASUREMENT ZONE/TARGET DISTANCE - The measurement zone is proportional to the distance
the RayTemp is away from the target. The RayTemp 8 is equipped with a 12:1 lens. If the target is
60 cm (24 ") away the measurement zone will be 5 cm (2 ") across.
EMISSIVITY - The RayTemp 8 has a default emissivity of 0.95 but is adjustable from 0.10 to 1.00.
The closer the emissivity value is set to the actual emissivity value of the object being measured,
the more accurate the readings will be. To adjust the emissivity press the MODE button so that 'E' is
display, press the button again so that ' ' is displayed. Use the UP and DOWN buttons to
set the emissivity required. Please note: non-contact infrared thermometers are not recommended
for use in measuring shiny or polished surfaces.
LASER POINTER - The circular laser function can be turned on or off by pressing the trigger and
simultaneously pressing the °C/°F button whilst the trigger is being pressed. The centre of the
measurement zone is 14 mm (approximately ½ ") above the lowest laser dot with the instrument
held in a horizontal position. For safety, the laser pointer will only activate when the trigger button
is pressed. The laser module is a Class 3 device that has a maximum power output of less than
5 mW at a wavelength of 660 nm. Prolonged, continuous exposure, such as staring at the beam,
can be harmful and should be avoided. DO NOT look at the beam of any optical instrument.
LCD BACKLIGHT - The LCD backlight function can be turned on or off by pressing the trigger and
simultaneously pressing the LOCK button whilst the trigger is being pressed. The backlight icon
' ' will be shown to indicate on or off.
UNDER RANGE/OVER RANGE - If the measured temperature goes below or above the range of the
instrument, 'Er' will be displayed. When the measured temperature returns within the range of the
instrument, readings will automatically be displayed.
THERMOCOUPLE PROBE/SENSORS - To display the thermocouple probe temperature, press the
MODE button until 'PRB' is displayed. If the thermocouple breaks or is not connected, the instrument
will display 'noP'. The RayTemp 8 should only be used with thermocouple type K nickel-chromium/
nickel-aluminium probes or sensors that meet BS EN 60584:1996 standard and are fitted with a
suitable miniature thermocouple plug. To review the max or min temperatures of the connected
probe whilst in PRB mode press the DOWN button for min and the UP button for max.
MAXIMUM, MINIMUM, DIFFERENTIAL & AVERAGE - To access the maximum reading, press the
MODE button until 'MAX' is displayed. To access the minimum reading, press the MODE button
again until 'MIN' is displayed. To access the differential reading, press the MODE button until 'dIF'
is displayed. To access the average reading, press the MODE button again until 'AVG' is displayed.
Maximum and minimum, differential and average readings will be reset when the instrument is
turned off.
Raytemp
®
8
Infrared Thermometer
814-045/26.03.15
www.etiltd.com
Operating Instructions
order code: 814-045
Question and Answer IconNeed help?

Do you have a question about the ETI Raytemp 8 and is the answer not in the manual?

Summary

Instrument Operation

Instrument Operation

Basic instructions for aiming and using the trigger to measure surface temperature.

Measurement Zone and Target Distance

Explains the relationship between distance and measurement zone size with a 12:1 lens.

Emissivity Adjustment

Details on adjusting emissivity from 0.10 to 1.00 for accurate readings.

Laser Pointer Function

Instructions for activating and safety precautions for the Class 3 laser pointer.

LCD Backlight

How to turn the LCD backlight on or off using the trigger and LOCK button.

Under;Over Range Indicators

Explains the 'Er' display for temperatures outside the instrument's range.

Thermocouple Probe;Sensors

Using thermocouple probes, error codes ('noP'), and viewing max/min temps.

Max, Min, Differential & Average

How to access and view maximum, minimum, differential, and average readings.

Temperature Scale and Alarms

Temperature Scale (°C;°F)

How to switch between Celsius and Fahrenheit scales.

High and Low Temperature Alarms

Setting and understanding high/low temperature alarm parameters and indicators.

Lock Function

Disables auto-power off for continuous monitoring up to 60 minutes.

Battery Replacement

Indicator for low battery and replacement instructions with AAA batteries.

Storage, Cleaning, and Safety

Storage and Cleaning

Guidelines for cleaning the sensor lens and storing the instrument.

EMC;RFI Considerations

Potential impact of high-frequency radio fields or electrostatic shock.

Guarantee Information

Details on the one-year guarantee and exclusions for defects or misuse.

Advanced Features and Maintenance

Temperature Scale and Alarms (FR)

Operating instructions for temperature scale and high/low alarms in French.

Lock Function (FR)

Instructions for the lock function to disable auto-power off in French.

Battery Replacement (FR)

Information on battery replacement and low battery indicators in French.

Storage and Cleaning (FR)

Guidelines for storing and cleaning the infrared thermometer in French.

EMC;RFI Considerations (FR)

Notes on EMC/RFI effects on instrument performance in French.

Device Operation and Settings

Device Operation (DE)

Instructions for operating the device and measuring surface temperature in German.

Measurement Range & Distance (DE)

Details on measurement range and target distance in German.

Emissivity Setting (DE)

Guidance on adjusting emissivity for accurate measurements in German.

Laser Pointer and Backlight (DE)

Information on laser pointer and LCD backlight functions in German.

Under;Over Range (DE)

Explanation of 'Er' display for out-of-range temperatures in German.

Thermoelement Sensors (DE)

Notes on using thermocouple sensors and error codes in German.

Max, Min, Differential & Average (DE)

How to access max, min, differential, and average readings in German.

Temperature Scale and Alarms (DE)

Switching scales and setting temperature alarms in German.

Lock Function and Batteries (DE)

Using the lock function and battery replacement in German.

Storage, Cleaning & EMC;RFI (DE)

Storage, cleaning, and EMC/RFI considerations in German.

ETI Raytemp 8 Specifications

General IconGeneral
Field of View8:1
Battery2 x AAA
Distance to Spot Ratio8:1
Auto Power OffYes, after 15 seconds
Weight180 grams
Resolution0.1 °C
Battery Life180 hours continuous use
Response TimeLess than 1 second
DisplayLCD
EmissivityAdjustable, 0.10 to 1.00