56
CALIB
calibration program
iii
‘S’ -> Save new calibration in KT-9
Pressing
s
will save the newly established calibra-
tion value in the KT-9’s memory for this particular
external sensor.
iv
‘N’ -> Next KT-9 for calibration
This option allows you calibrate another KT-9.
v
‘E’ -> Exit program & DISABLE COM1 output
Pressing
e
will stop the program, close the com-
munications link and return full manual control to
the KT-9.
The particular external sensor is now calibrated and ready to
use. This calibration remains in the KT-9s memory until a new
calibration for a different sensor is performed. When you
disconnect the external sensor, the KT-9 automatically switches
back to its internal sensor but remembers the external sensor’s
calibration for use when it is reconnected.
When switching to a different external sensor, the calibra-
tion routine will have to be re-run every time.
The KT-9 uses flash memory. Consequently, it will not lose
the calibration value (or any data) when the batteries are
removed.
56
CALIB
calibration program
iii
‘S’ -> Save new calibration in KT-9
Pressing
s
will save the newly established calibra-
tion value in the KT-9’s memory for this particular
external sensor.
iv
‘N’ -> Next KT-9 for calibration
This option allows you calibrate another KT-9.
v
‘E’ -> Exit program & DISABLE COM1 output
Pressing
e
will stop the program, close the com-
munications link and return full manual control to
the KT-9.
The particular external sensor is now calibrated and ready to
use. This calibration remains in the KT-9’s memory until a new
calibration for a different sensor is performed. When you
disconnect the external sensor, the KT-9 automatically switches
back to it’s internal sensor but remembers the external sensor’s
calibration for use when it is reconnected.
When switching to a different external sensor, the calibra-
tion routine will have to be re-run every time.
The KT-9 uses flash memory. Consequently, it will not lose
the calibration value (or any data) when the batteries are
removed.