Data accuracy
73
Data accuracy
Susceptibility
KT-5 KT-9
Use the N
O
-P
IN
mode to
obtain the highest sample
data accuracy (“P
IN
vs. N
O
-P
IN
modes — comments” on
page 74).
KT-5 KT-9
The KT-5 was designed to
measure the A
PPARENT
suscep-
tibility.
At low susceptibility values,
the differences between
A
PPARENT
and T
RUE
are rela-
tively low.
Above values of 100 x 10
-3
SI
units, you must correct the
readings to T
RUE
. The correc-
tion data and procedure is
shown in Figure 15 on page
13 of the KT-5 manual.
The KT-9 automatically com-
putes the T
RUE
susceptibility.
Data accuracy
73
Data accuracy
Susceptibility
KT-5 KT-9
Use the N
O
-P
IN
mode to
obtain the highest sample
data accuracy (“P
IN
vs. N
O
-P
IN
modes — comments” on
page 74).
KT-5 KT-9
The KT-5 was designed to
measure the A
PPARENT
suscep-
tibility.
At low susceptibility values,
the differences between
A
PPARENT
and T
RUE
are rela-
tively low.
Above values of 100 x 10
-3
SI
units, you must correct the
readings to T
RUE
. The correc-
tion data and procedure is
shown in Figure 15 on page
13 of the KT-5 manual.
The KT-9 automatically com-
putes the T
RUE
susceptibility.