421502-en-GB_V2.0 3/19
Specifications
5-digit multifunction LCD
9V (NEDA 1604, IEC 6F22, or 006P)
Meter switches OFF after approx. 30 mins. of inactivity
160 hours typical with carbon zinc battery
192 x 91 x 52.5mm (7.5 x 3.6 x 2.1”) / 365g (11.7 oz.)
Type-J thermocouple range
-200 ~ 1050
o
C (-328 ~ 1922
o
F)*
Type-K thermocouple range
-200 ~ 1370
o
C (-328 ~ 2498
o
F)*
*Safety note: The supplied type-K thermocouple can only be used
to measure temperature to 260
o
C (500
o
F). To measure higher
temperatures please obtain a thermocouple rated for higher
temperature
± (0.05% rdg + 0.3
o
C) -50
o
C ~ 1370
o
C
± (0.05% rdg + 0.7
o
C) -50
o
C ~ -200
o
C
± (0.05% rdg + 0.6
o
F) -58
o
F ~ 2498
o
F
± (0.05% rdg + 1.4
o
F) -58
o
F ~ -328
o
F
0.1 times the applicable accuracy specifications per
o
C from 0
o
C ~
18
o
C and 28
o
C ~ 50
o
C (32
o
F ~ 64
o
F and 82
o
F ~ 122
o
F)
Gasket protected front panel
24VDC or 24VAC rms max. input on any combination of inputs
Accepts standard miniature thermocouple connectors
Type-K with Teflon® insulation. Cable length: 4 ft. (1.2m)
Max. insulation temperature: 260
o
C (500
o
F)*
Accuracy: ± 2.2
o
C (4
o
F) or ± 0.75% of reading (whichever is greater)
*Safety note: The supplied thermocouple can only be used to
measure temperature to 260
o
C (500
o
F). To measure higher
temperatures please obtain a thermocouple rated for higher
temperature.
0 to 50
o
C (32 to 122
o
F); less than 80% RH
-20 to 60