Section 7: Technical Specifications
This section presents the specifications and safety standards of the product.
NOTE: The following specifications are subject to change and are only
noted as a point of reference.
Product name: Fetal Doppler
Mode: JPD-100B
Safety: Complies with IEC 60601-1, EN 60601-1, EN60601-1-2, IEC 61266
Physical Dimensions:
129mm W x 83mm D x 34 mm H, Wt: 0.24kg (not including battery)
Classification:
Anti-electric Shock Type: Internally powered equipment
Anti-electric Shock Degree: Type B applied part
Degree of Protection against Harmful Ingress of Water:
Main Unit: Non-protected
Probe: IPX4 Water Ingress Protection Code
Degree of Safety IN Presence of Flammable Gases: Equipment not suitable
for use in presence of flammable gases
Working Mode: Continuous operation
Technical parameters:
Ultrasound:
Ultrasonic emitting frequency: 3.0MHz
Overall sensitivity at the distances 200mm from the face of the probe
(Doppler frequency: 500±50Hz, Target velocity: 10cm/s~40cm/s): 92.4dB
Spatial-peak temporal-peak acoustic pressure: 20.1kPa
Output power: 7.9mW