CISPR 11
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions
IEC
61000-3-2
Class A
Voltage
fluctuations
/flicker
emissions
IEC
61000-3-3
Not
applicable
Guidance & Declaration—electromagnetic immunity
The models JPD-100B are intended for use in the electromagnetic
environment specified below. The customer or the user of the models
JPD-100B should assure that it is used in such an environment.
IEC 60601
test level
Compliance
level
Electromagnetic
environment-
guidance
±6 kV
contact
±8 kV air
±6 kV
contact
±8 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30 %.
±2kV for
power supply
lines
±1kV for
Input/output
lines
±2kV for
power supply
lines
Mains power quality
should be that of
atypical commercial
or hospital
environment
±1kV line to
line
±2kV line to
earth
±2kV line to
earth
Mains power quality
should be that of a
typical commercial
or hospital
environment.