Measuring Temperature
With probes (adapter and probes are optional)
Connect the thermocouple adapter to the two input terminals of the clamping ;
make sure polarity is respected.
(- on the COM terminal and + on the + terminal)
Connect the desired probe to the adapter.
712-01 Adapter
DX.12.06 Flexible probe
DX.12.08 Bracelet probe
DX.12R Extension (is used with the bracelet probe)
DX.12.11 Standard needle probe
DX.12.15 Surface contact probe
DX.12.17 Air probe
The T° EXT symbols appear on the display.
DX.12-11
DX.12-15
DX.12-17
DX.12-06
DX.12-08
DX.12-11
DX.12R
DX.12-15
DX.12-17
712-01
DX.12-06
DX.12-08
Internal Temperature External Temperature External Temperature
Probe Type Integrated Circuit K Couple K Couple
Display Range ° C ° C ° C
Measurement -10° C to + 50° C -50.0° C to + 399,9° C +400° C to + 1,000° C
Range
+15° F to +120° F -50° F to + 399,9° F +400° F to + 1,800° F
Precision ± 1,5° C 1% L ± 1,5° C 1% L ± 1,5° C
± 2,7° F 1% L ± 2,7° F 1% L ± 2,7° F
Resolution 0,1° C 0,1° C 1° C
0,1° F 0,1° F 1° F