3. Product specifications
Digital Clinical Thermometer
d.c. 3.0V, one CR1632 battery
Adjusted mode / Direct mode
32.00
℃
~ 42.90
℃
(89.60
℉
~109.22
℉
)
±0.10℃: between 35.00℃~42.00℃
(±0.18℉ between 95.00℉~107.60℉)
±0.20℃(or±0.36℉): outside this ranges
Adjusted mode:Less than 8s (for water bath)
Direct mode:Less than 50s (for water bath)
High body temperature hint
Temperature:5℃~40℃ (41℉~104℉)
Relative humidity:15%~95%RH,No condensing
Atmospheric pressure:70 kPa~106 kPa
Transportation and storage
Temperature:-25℃~55℃ (-13℉~131℉)
Relative humidity:15%~95%RH,No condensing
Atmospheric pressure:70 kPa~106 kPa
5 years (excluding vulnerable parts)
Internally powered ME equipment
Type BF applied part,including the whole unit.
*Statement: If stored or used outside the temperature, humidity or atmospheric pressure range specified by
the manufacturer, the product may not achieve the claimed performance.