6 ENGLISH
OXYGEN/PRESSURE PORT CONNECTOR
If pressure readings and/or additional oxygen are required, an Oxygen/Pressure Port Connector
is available (REF 900HC452). Please contact your healthcare provider.
TECHNICAL SPECIFICATIONS
F&P ESON 2 EXHAUST FLOW
The F&P Eson 2 Nasal Mask has a unique Diuser System that is designed to diuse air which
is expelled from the mask. This reduces draft. The Diuser does not significantly change the
amount of air flowing from the mask; therefore, the mask can be used with or without the
Diuser.
Pressure (cmH
2
O) 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Flow with Diuser (L/min) 19 25 29 34 38 41 45 48 52 55 58 61 63 66
Flow without Diuser (L/min) 27 34 39 44 48 52 56 59 62 66 69 71 74 77
Due to manufacturing variations, the exhaust flow rates may vary from the nominal values
shown above.
Exhaust Flow (L/min)
Pressure (cmH
2
O)
WITHOUT
Diuser
WITH
Diuser
80
60
40
20
0
4 6 8 10 12 14 16 18 20 22 24 26 28
30
RESISTANCE TO FLOW
Pressure drop through mask with Diuser @ 50 L/min: 0.19 ± 15% cmH
2
O
Pressure drop through mask with Diuser @ 100 L/min: 0.77 ± 15% cmH
2
O
Pressure drop through mask without Diuser @ 50 L/min: 0.17 ± 15% cmH
2
O
Pressure drop through mask without Diuser @ 100 L/min: 0.71 ± 15% cmH
2
O
MASK DEAD SPACE
Small: 69 cc Medium: 86 cc Large: 98 cc
SOUND
The sound power level of mask with Diuser is 21.3 dBA, with uncertainty 2.5 dBA
The sound pressure level of the mask with Diuser is 13.3 dBA, with uncertainty 2.5 dBA.
The sound power level of mask without Diuser is 31.4 dBA, with uncertainty 2.5 dBA
The sound pressure level of the mask without Diuser is 23.4 dBA, with uncertainty 2.5 dBA.
The F&P Eson 2 Nasal Mask and accessories are not made with natural rubber latex.