6 ENGLISH
OXYGEN/PRESSURE PORT CONNECTOR
If pressure readings and/or additional oxygen are required, an Oxygen/Pressure Port Connector
is available (REF 900HC452). Please contact your healthcare provider.
TECHNICAL SPECIFICATIONS
F&P ESON 2 EXHAUST FLOW
The F&P Eson 2 Nasal Mask has a unique Diuser System that is designed to diuse air which
is expelled from the mask. This reduces draft. The Diuser does not significantly change the
amount of air flowing from the mask; therefore, the mask can be used with or without the
Diuser.
Pressure (cmH
2
O) 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Flow with Diuser (L/min) 19 25 29 34 38 41 45 48 52 55 58 61 63 66
Flow without Diuser (L/min) 27 34 39 44 48 52 56 59 62 66 69 71 74 77
Due to manufacturing variations, the exhaust flow rates may vary from the nominal values
shown above.
Exhaust Flow (L/min)
Pressure (cmH
2
O)
WITHOUT
Diuser
WITH
Diuser
60
40
20
0
4 6 8 10 12 14 16 18 20 22 24 26 28
RESISTANCE TO FLOW
Pressure drop through mask with Diuser @ 50 L/min: 0.19 ± 15% cmH
2
O
Pressure drop through mask with Diuser @ 100 L/min: 0.77 ± 15% cmH
2
O
Pressure drop through mask without Diuser @ 50 L/min: 0.17 ± 15% cmH
2
O
Pressure drop through mask without Diuser @ 100 L/min: 0.71 ± 15% cmH
2
O
MASK DEAD SPACE
Small: 69 cc Medium: 86 cc Large: 98 cc
SOUND
The sound power level of mask with Diuser is 21.3 dBA, with uncertainty 2.5 dBA
The sound pressure level of the mask with Diuser is 13.3 dBA, with uncertainty 2.5 dBA.
The sound power level of mask without Diuser is 31.4 dBA, with uncertainty 2.5 dBA
The sound pressure level of the mask without Diuser is 23.4 dBA, with uncertainty 2.5 dBA.
The F&P Eson 2 Nasal Mask and accessories are not made with natural rubber latex.
For single patient reuse, the mask should be replaced no later than 12 months after the date of
first use.