Finometer® Model-2
User's Guide
Finometer
® Midi User’s Guide
This document is for information purposes only
v01.04
10-2012
© Finapres Medical Systems B.V.
0344
Ref. nr: 901700-584-53

Table of Contents

Questions and Answers

  • S
    Sarah GrayAug 13, 2025
    What to do if the FINAPRES Finometer Midi Medical Equipment shows 'CHECK AIR SUPPLY'?
    • D
      David MurrayAug 13, 2025
      If the FINAPRES Medical Equipment displays 'CHECK AIR SUPPLY', it indicates the pump system isn't generating enough pressure or flow. First, ensure the finger cuff air hose is correctly inserted into the Frontend Unit and that the air hose between the Frontend Unit and the main unit is properly connected. Check both hoses for any leaks or kinks. If the error continues, it is recommended to have the device inspected by Finapres Medical Systems B.V.
  • J
    Judy GarciaAug 16, 2025
    What to do if FINAPRES Medical Equipment shows 'UNACCEPTABLE P/V'?
    • S
      sarahleeAug 17, 2025
      If the FINAPRES Medical Equipment displays 'UNACCEPTABLE P/V', it signifies a pressure or plethysmogram error. This occurs if the pressure has been below 10 mmHg or above 250 mmHg for 2.5 seconds, or if the plethysmogram wasn't sufficiently small during the measurement. Try again. Ensure the finger cuff is wrapped tightly enough around the finger. Also, check for issues related to 'CHECK AIR SUPPLY' and 'CONNECT CUFF CABL' errors.
  • D
    Dana HoffmanAug 20, 2025
    Why does my FINAPRES Finometer Midi Medical Equipment show 'NO PLETHYSMOGRAM'?
    • K
      Kimberly JacobsonAug 20, 2025
      The 'NO PLETHYSMOGRAM' error on the FINAPRES Medical Equipment is usually caused by contracted arteries. To resolve this, retry the measurement. You might also try using another finger or warming the hand to improve circulation.
  • T
    timothy56Aug 25, 2025
    What should I do if FINAPRES Medical Equipment displays 'PWR SUP NOT WELL'?
    • L
      Laura MurrayAug 25, 2025
      If the FINAPRES Medical Equipment shows 'PWR SUP NOT WELL', it means a short-circuit, zero offset, or power supply unbalance has occurred. Immediately turn off the device. Briefly retry powering it on without the finger cuff attached. This error can also appear if you disconnect the Frontend Unit during a measurement. In this case, reconnect the Frontend Unit and power the unit off and on again. If the problem persists, the device should be checked by Finapres Medical Systems B.V.
  • J
    John FuentesAug 28, 2025
    What causes 'UNSTABLE PRES' error on FINAPRES Finometer Midi, and how to fix it?
    • N
      nicholasfrancisAug 28, 2025
      The 'UNSTABLE PRES' error on the FINAPRES Medical Equipment indicates that constant pressure levels were not stable enough during the pressure staircase procedure at start-up or during a physiocal. Check if the finger cuff air hose is properly inserted into the Frontend Unit. Then, retry the procedure. If the error continues, it is recommended to have the device inspected by Finapres Medical Systems B.V.
  • N
    Nicole Montgomery MDSep 1, 2025
    How to resolve 'FINGER TOO THIN' error on FINAPRES Finometer Midi?
    • K
      Keith NewtonSep 1, 2025
      The 'FINGER TOO THIN' error on the FINAPRES Medical Equipment means the infrared level transmitted through the finger is too high. First, retry the measurement. If the issue persists, try using a smaller finger cuff. If that doesn't work, wrap some thin cloth around the finger before replacing the cuff. Use darker cloth if necessary.
  • V
    Vicki FosterSep 4, 2025
    What does 'INT GND FAULT' mean on my FINAPRES Finometer Midi?
    • N
      Noah LewisSep 4, 2025
      The 'INT GND FAULT' error on the FINAPRES Medical Equipment indicates that the zero offset of the pressure transducer is too large. You can try to resolve this by powering the device off and on a number of times. If the error continues, it is recommended to have the device inspected by Finapres Medical Systems B.V.
  • J
    jacksonrebeccaSep 8, 2025
    What does 'A/D CNV NOT WELL' mean on FINAPRES Finometer Midi?
    • C
      Cindy JensenSep 8, 2025
      The 'A/D CNV NOT WELL' error on the FINAPRES Medical Equipment indicates that the signal ground voltage of the Analog/Digital converter is too high. Power the device off and then on again. If the error persists, have the device checked by Finapres Medical Systems B.V.

Summary

The Finometer Model-2 Components

BeatScope Easy Software

PC software for monitoring, recording, and remote control of the device.

Cautions in Using Finometer Model-2

General Usage Cautions

Important warnings and precautions for safe and effective device operation.

Avoiding Injury to Subjects and Personnel

Safety Precautions

Essential safety measures to prevent harm to users and subjects.

1 Introduction

What is Finometer Model-2?

Overview of the device, its measurement method, and key features.

1.3 Quickstart

Initial Setup Steps

Step-by-step instructions for setting up and preparing the device for use.

1.4 A First Finometer Model-2 Measurement

Performing First Measurement

Guide on how to conduct the initial blood pressure measurement.

2 Safety Information

Warnings and Subject Safety

Critical warnings related to device usage and ensuring subject safety.

Cautions

Important considerations for preventing device damage and ensuring proper function.

Precautions

Measures to ensure measurement accuracy and prevent operational issues.

Protective Measures

Safety mechanisms and features built into the device.

3 System Description

Finometer Model-2 Control

How to operate the device using the control keypad and LCD display.

Connecting the Frontend Unit

Instructions for connecting the frontend unit to the subject and main unit.

4 Using the Control Keypad

Error Mode

The mode indicating an error condition and requiring operator intervention.

5 Configuring the Device

Height Correction Unit Nulling Procedure

Procedure to calibrate the HCU by setting its zero offset.

PC Monitoring and Recording

Steps for setting up PC connectivity for data monitoring and recording.

6 Performing a Measurement

Using Finger Cuffs

Comprehensive guide on finger cuff selection, handling, and application.

Finger Cuff Selection

How to choose the correct finger cuff size for optimal measurement.

Cuff Application

Step-by-step instructions for correctly applying the finger cuff.

Starting a Measurement

Procedure for initiating a blood pressure measurement cycle.

7 Error Messages and Troubleshooting

Error Messages

A comprehensive list of error messages and recommended actions.

Plethysmogram;Cuff Errors

Errors related to finger signal detection and cuff issues.

Air Pressure;Cuff Errors

Errors concerning air supply, pressure, and cuff connectivity.

Missing Sub Device Errors

Errors indicating a disconnected or missing component.

Fatal Internal Errors

Critical internal errors requiring professional service.

8 Maintenance, Storage, and Calibration

FINAPRES Finometer Midi Specifications

General IconGeneral
BrandFINAPRES
ModelFinometer Midi
CategoryMedical Equipment
LanguageEnglish