2120
Safety & Compliances
Maximum voltage between any terminal
and earth ground
: 1000 V ac/dc
Compliances : Complies with CSA C22.2 No 1010.1-92,
ANSI/ISA-S82, 01-94 to 1000 V Overvoltage
Category lll.
Certifications : UL & cUL standard UL 3111-1 Listed
CE-marking certificated
Surge Protection : 8 kV peak per IEC 1010.1-92
Fuse Protection for mA or µA inputs : 1000 V / 440 mA lR 10 kA FAST fuse
Fuse Protection for A input : 1000 V / 11 A lR 10 kA FAST fuse
Physical Specifications
Display (LCD) : Digital – 4000 counts display;updates
5 times/sec.
Operating Temperature : 0 C to 40 C
Storage Temperature : –20 C to 60 C
Temperature Coefficient : nominal 0.15 x (specified accuracy) / C
@(0 C to 18 C or 28 C to 40 C),
or otherwise specified
Relative Humidity : 0 % to 80 % @ (0 C to 35 C)
0 % to 70 % @ (35 C to 40 C)
Altitude : Operating – up to 2000m
Storage – 10000m
Battery Type : Single 9V battery –NEDA 1604, JIS 006P or
IEC 6F 22
7. SPECIFICATIONS
Battery Life : 250 hrs. typical (with backlight off) [703]
750 hrs. typical (with backlight off) [701]
Shock Vibration : Per MIL-T-PRF 28800 for Class II instruments
Pollution Degree : 2
Electromagnetic
Compatibility (EMC) : Susceptibility – Commercial Limits for
EN 50082-1
Emissions – Commercial Limits for EN 50081-1
Size (H x W x L) : 40.5 x 92 x 172 mm
Weight : Approx. 386g
Warranty :
3 years
Calibration Interval : 1 year
Feature Summary
Backlight : For clear readings in poorly lighted areas
Fast Autoranging : Meter automatically selects the best range
momentarily
HOLD
:
Holds readings on display
Continuity / Open test : Beeper sounds
Battery/Fuse Access Door : battery or fuse replaceable without voiding
calibration
High-Impact Overmolded Case : Protective holster features