FISCHERSCOPE
®
X-RAY 189
Cleaning and Maintenance
Instead of a measurement
reading, the window Open
appears after a measure-
ment has been started
The communication between
instrument and WinFTM is
disabled.
Enable the communication be-
tween instrument and WinFTM
(see Chap. 3.7 on Page 27).
Communication between in-
strument and WinFTM can-
not be established
The instrument is not correct-
ly connected to the comput-
er.
Check and correct the connection
between instrument and computer
(see Chap. 3.7 on Page 27),
switch the instrument off and on.
A menu or menu option
does not appear in the menu
bar
The short menu is turned on. Turn on the long menu by selecting
General > full menu (requires
password).
Menu options to edit Def.MA
files are grayed out and
cannot be selected
The WinFTM extension SU-
PER is not active.
If you already have the WinFTM
extension SUPER, shut down Win-
FTM, plug the SUPER FISIM dongle
and start WinFTM again.
If you not yet own the WinFTM ex-
tension SUPER, contact your FIS-
CHER representative.
XY(Z) stage cannot be mo-
ved (XDLM 237)
The XY(Z) stage has been
moved up, down, right or
left too far.
Move the XY(Z) stage in the oppo-
site direction.
The safety system is en-
gaged, e. g. the stage has
been jammed by a foreign
object.
Remove the foreign object.