EasyManua.ls Logo

Fisher AD-530 - Adjustment Procedures; Pre-Adjustment Checks & Settings; Focus, Tracking, and Diffraction Adjustments

Fisher AD-530
19 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ADJUSTMENT
PROCEDURES
BEFORE
CHECKING
OR
ADJUSTING
CD
PLAYER
1.
Procedures
for
all
adjustments
for
the
CD
player
from
start
to
finish
are
described
below.
2.
The
pick-up
does
not
have
to
be
adjusted
if
everything
has
been
checked
and
no
problems
were
found.
CHECKING
FOCUS
ERROR
SIGNAL
The
effectiveness
of
the
pick-up
is
determined
by
checking
its
focus
error
signal.
This
must
be
checked
before
any
repairs
are
performed
on
the
CD
player.
1...
Apply
Power
Switch
to
the
set
and
press
the
STOP
button.
2.
Cut
Main
P.C.Board
jumper
cable
J223
and
connect
an
oscilloscope
between
P202
left
terminal
and
the
GND.
(Set
the
time
at
0.5Sec/DIV.)
3.
Press
the
OPEN/CLOSE
button,
place
the
test
disc
on
the
table
load,
and
then
press
the
OPEN/CLOSE
button
once
again.
:
4.
Once
the
disc
is
set,
a
waveform
like
that
shown
in
the
illustration
below
will
be
repeated
three
times.
Check
that
the
p-p
voltage
of
this
waveform
is
the
same
above
and
below
the
horizontal
axis
at
3V.
Repeated
three
times
*
p-p
voltage
of
3V
*
+
and
voltage
roughly
same
5.
Press
the
OPEN/CLOSE
button
repeatedly
to
pull
out
and
insert
the
disc
until
the
waveform
appears
each
time
the
disc
is
inserted
and
removed.
6.
If
the
results
of
the
above
procedure
are
not
satisfactory,
the
pick-up
is
damaged.
7.
After
completing
checks
of
the
focus
error
signal
solder
the
jumper
line
(J223)
back
in
place.
FREE
RUN
FREQUENCY
ADJUSTMENT
1.
Connect
the
frequency
counter
to
VCO
test
terminal
(CN11-1)
PCK
test
pin
and
to
GND.
(Use
a
10:1
probe.)
2.
Shortthe
test
point
TP
DSP
OFF
pin
to
GND.
3.
Apply
Power
Switch
to
the
set
and
press
the
STOP
button.
4.
Adjust
the
OSC
coil
T202
using
a
plastic
screwdriver
until
the
frequency
counter
indicator
reads
4.3218MHz.
SETTING
OF
INITIAL
POSITION
OF
VOLUME
1.
Set
the
variable
resistors
to
the
initial
positions
listed
below:
P202
(FOCUS
GAIN)
.........eeeeeeeeeeees
Mechanical
Center
P203
(E/F
BALANCE)
..........ccseeeeees
Mechanical
Center
P204
(TRACKING
GAIN)
............66
Mechanical
Center
P205
(TRACKING
OFFSET)
...........
Mechanical
Center
P206
(KICK
GAIN)
....----sesseeeeeeeen
sees
Mechanical
Center
FOCUS
OFFSET
ADJUSTMENT
1.
Apply
Power
Switch
to
the
set
and
press
the
STOP
button.
2.
Connect
a
DC
Voltmeter
to
the
test
terminal
(CN9-5)
F
(focus
coil)
pin
and
short
the
focus
test
terminal
(CN10-2)
FOF
pin
to
GND.
3. In
this
condition
check
that
the
DC
Voltmeter
indicator
is
in
the
range
—0.7V
~
—1.2V.
TRACKING
OFFSET
ADJUSTMENT
1.
Apply
Power
Switch
to
the
set
and
press
the
STOP
button.
2.
Connect
a
DC
Voltmeter
to
the
test
terminal
(CN9-3)
T
(tracking
coil)
pin
and
short
the
focus
test
terminal
TOF
pin
to
GND.
3.
‘In
this
condition
check
that
the
DC
Voltmeter
indicator
is
in
the
range
OV+200mV,
and
adjust
P205
if
it
is
not.
NOTE:
Perform
this
adjustment
once
again
after
adjusting
the
E/F
BALANCE
and
TRACKING
GAIN
signals.
DIFFRACTION
GRID
ADJUSTMENT
This
set
uses
a
three-beam
laser
pick-up
system.
Adjusting
the
diffraction
grid
means
adjustment
so
that
beams
E
and
F
are
in
position
with
respect
to
the
main
beam
as
shown
in
the
diagram
below
on
the
bit
streams.
The
adjustment
procedure
involves
arranging
the
three
beams
in
a
straight
line
along
the
same
bit
stream
and
then
adjusting
them
to
specified
positions.
The
adjustment
steps
are
detailed
below.
These
are
very
fine
adjustments
so
great
care
should
be
exercised
in
this.
DIFFRACTION
GRID
ADJUSTMENT
1.
Apply
Power
Switch
to
the
set
and
press
the
STOP
button.
2.
Short
tester
terminal
(CN9-1,2)
microprocessor
TEST
pin
to
the
TEST
pin,
and
then
short
test
terminal
(CN12-1)
TOFF
pin
to
GND
to
turn
the
tracking
servo
OFF.
3.
In
this
state,
connect
the
E-beam
output
of
test
terminal
(CN12-6)
to
Channel
1
of
the
oscilloscope
and
F-beam
output
of
test
terminal
(CN12-4)
to
Channel
2,
and
now
observe
the
lissajous
waveform
on
the
oscilloscope.