16
17
In accordance with Directive 2012/19/EC, the symbol on the equipment indicates that the equipment to be disposed of (excluding
accessories) is considered as waste and must therefore be subject to 'separate collection'. Therefore, the user must deliver this waste
(or have it delivered) to the separate collection centres set up by the local authorities, or hand it over to the retailer when purchasing a
new appliance of an equivalent type. Separate waste collection and subsequent treatment, recovery and disposal operations promote
the production of equipment from recycled materials and limit the negative environmental and health eects caused by improper waste
management. Unauthorised disposal of the product by the user entails the application of the administrative sanctions provided for in
the laws transposing Directive 2012/19/EC of the member state or country where the product is disposed of.
Accessories (Secretion collection vessel and piping)
Accessories to be disposed of as general waste after a sanitisation cycle.
Packaging
20
PAP
Product box
04
LDPE
Product packaging bag
DISPOSAL
Aspiration unit
TROUBLESHOOTING
Before performing any operation, switch o the appliance and unplug the mains cable from the socket.
PROBLEM CAUSE REMEDY
The device is not working
Power cable not correctly plugged into the
appliance socket or in the mains power
socket
- Plug the power cable correctly into the sockets
No aspiration
Collection vessel lid not properly inserted
in the vessel
- Insert the lid correctly into the collection vessel
Aspiration failure caused by
uid spillage
Clogged lter - Replace the lter
Blocked oat Fouling on the oat
- Remove the lid from the vessel, dismantle the pro-
tective device and remove the oat. Then carry out
cleaning operations as per the paragraph “HYGIENIC
PREPARATION”
Low and/or no vacuum power
Vacuum regulator all open
- Close the regulator completely and check vacuum
power
Protection lter blocked - Replace the lter
Obstructed, kinked or disconnected lter
and device connection tubes
- Check the condition of the tubes, replace them if
obstructed and connect them correctly as per the
“CONNECTION DIAGRAM” in this manual
Vessel lid overow valve closed or blocked - Unlock the overow valve, hold the device upright
Dirty or obstructed or damaged pump
- Take the device to your dealer or an authorised
FLAEM service centre
If, after checking the conditions described above, the device still is not working properly, we recommend that you contact your
trusted dealer or an authorised FLAEM service centre nearest to you. You can nd a list of all Service Centres at http://www.aemn-
uova.it/it/info/assistenza